8.Replace the front panel and fix it with 4 screws.

TEMPERATURE RECORDER (OPTION)

6.Attach the sensor cover to the right side of the chamber with 2 screws on the chamber wall (Fig. 4)

7.Seal the access port on the outside with putty as shown in Fig. 5.

8.Replace the front panel and fix it with 4 screws.

9.Operate the freezer until the chamber temperature gets to the set temperature. Check the recorded temperature and chamber temperature displayed on the control panel. Adjust the zero adjustment volume on the temperature recorder so that the recorded temperature can corresponds with the displayed temperature if they are not compliance each other.

Refer to the instruction manual enclosed with the recorder.

Putty

Fig. 4

Fig. 5