Connection on PCB

Connection on PCB

Following is the explanation of connections on Main PCB.

Connector

Connects to

Usage

Voltage

 

 

 

 

CN1

Switching power supply

To supply the power to PCB

#1:12V

 

 

 

 

CN2

MTR-480Option

To connect with interface

 

 

 

 

 

 

Remote alarm terminal

 

 

CN3

#1: COM.

Output for remote alarm contact

 

#2: N.O.

 

 

 

 

 

#3: N.C.

 

 

 

#1-#2: Temp. control relay

To control chamber temperature

#1-#2:12V

CN4

To conduct current to cap. tube

#5-#6:12V

#5-#6: Fan relay

 

heater

 

 

 

 

CN5

#1, #7: Buzzer

To control with each switches

 

#2-#6: Control PCB (CN51)

 

 

 

 

 

 

 

 

CN6

Display PCBCN52

To connect with each LED

 

 

 

 

 

 

 

To control condensing fan motor

 

CN7

#7-#8: Comp. sensor

by the temperature in comp.

 

 

 

sensor

 

CN8

#1-#2: Battery

To supply the power to Main PCB

#2:6V

during power failure

 

 

 

 

 

 

 

 

CN9

Unused

 

 

 

 

 

 

CN11

#1-#3: Temp. sensor

To detect chamber temperature

 

 

 

 

 