SPECIFICATIONS
Name | |||
Model |
| ||
Outer dimensions | W286 x D87 x H520 mm | ||
Applicable model | |||
Applied gas | Liquid CO2 gas | ||
Temperature control | Gas type thermostat | ||
Door switch | Door | ||
Indicator | |||
Accessories | Mounting kit (except for gas cylinder) |
| Mounting kit (except for gas cylinder) |
| Tank connecting tube: mm size |
| Tank connecting tube: inch size |
Power source | AC 115 V / AC 230 V (supplied from a freezer) | ||
Battery | DC 24 V, 4 Ah, Automatic recharge | ||
Weight |
| 11 kg | |
Optional component | Mounting plate |
Note : Design or specifications will be subject to change without notice.
The battery for power failure alarm is an article for consumption.
It is recommended that the battery will be replaced about every 3 years. Contact Sanyo sales representation or agent at the time of replacement of the battery for recycling.
PERFORMANCE
Temperature range | ||
Duration | Approx. 13 hours (liquid CO2 gas cylinder of 30 kg) | |
(ambient temp.; 30oC, set temp.; | ||
| ||
Maximum power consumption | 6 W | |
Usable environment | Temperature; |