ŶControl specifications
Item |
| Specifications | ||
|
|
| ||
Cooling performance | ||||
Temperature control range | ||||
Temperature controller | Microprocessor controlled system: | |||
|
| Temperature set range: | ||
|
| Set value kept in | ||
Temperature sensor |
| Pt.1000Ω | ||
Temperature display | LED digital display(1℃ increment) | |||
| Temperature | Selectable +5℃~+20℃ for high temp. alarm (Initial; +10℃) | ||
|
| ALARM lamp blinks and intermittent buzzer beeps with | ||
|
| 15min.of delay |
| |
|
| Remote alarm contact: Normal Open, rating 30VDC, 2A | ||
|
| Temperature alarm turns on during power failure | ||
|
| (not linked with buzzer) | ||
Alarm | Filter | None |
| |
Power failure | ALARM lamp blinks with intermittent buzzer beeps, remote | |||
| ||||
|
| alarm outputs. |
| |
| Remote alarm | 3P remote alarm terminal: Maximum 30VDC, 2A | ||
|
| |||
|
| Outputs during temperature alarm and power failure alarm. | ||
| Battery | BATTERY lamp is lit. | ||
|
| Accumulating period: 2.8 years | ||
Control panel | Lamp: ALARM, BATTERY | |||
|
| BUZZER: Buzzer key | ||
|
| ALARM TEST: Alarm test key | ||
|
| SET: To switch SV and PV | ||
|
| >: To shift the digit | ||
|
| ȁ: To change the digit | ||
Key lock function | With > key is pressed for 5 seconds; | |||
|
| L0: Unlocked | L1: Locked | |
Self diagnosis function | When any failure occurs on the temperature sensor, | |||
|
| ・Error code and internal temp.(PV) are alternately displayed. | ||
|
| ・Remote alarm contact is ON and the buzzer beeps. | ||
Power switch | None |
| ||
Compressor protection | Overload relay, Comp. sensor |
3