F25: | <Purpose> | Alarm auto recovery time setting |
|
|
|
|
| <Operation> | Input F25 and press SET key to display “030” (initial). | ||||
|
| Change to the desired value with | key and |
| key. | |
|
| Press SET key to store the value and revert to PV display. | ||||
|
| 000: Auto recovery OFF | 040: Recovers after 40min. passed | |||
|
| 010: Recovers after 10min. passed | 050: Recovers after 50min. passed | |||
|
| 020: Recovers after 20min. passed | 060: Recovers after 60min. passed | |||
|
| 030: Recovers after 30min. passed |
|
|
|
|
|
| (initial) |
|
|
|
|
F28: | <Purpose> | To display delay time of permission of measuring running rate | ||||
|
| (2hrs timer; 000~120 min) |
|
|
|
|
| <Operation> | Service code should be input in F06 prior to use this mode. | ||||
|
| Input F28 and press SET key to display F28 and ‘xxx’ (present count | ||||
|
| value) alternately. Press SET key to revert to PV. | ||||
F29: | <Purpose> | To display delay time of permission of measuring running rate | ||||
|
| (8hrs timer; 000~480 min) |
|
|
|
|
| <Operation> | Service code should be input in F06 prior to use this mode. | ||||
|
| Input F29 and press SET key to display F29 and ‘xxx’ (present count | ||||
|
| value) alternately. Press SET key to revert to PV. | ||||
F30: | <Purpose> | ROM version is displayed |
|
|
|
|
| <Operation> | Service code should be input in F06 prior to use this mode. | ||||
|
| Input F30 and press SET key to display alternately F30 with “X.XX” | ||||
|
| (present ROM version). |
|
|
|
|
|
| Press SET key to revert to PV display. |
|
| ||
F31: | <Purpose> | Simply passing through if entered by mistake. | ||||
| <Operation> | Press SET key in “F31” displayed to revert to PV display. | ||||
|
| Buzzer tone is emitted if service code is not input in F06. | ||||
F32: | <Purpose> | To notify accumulation time of condensing fan motor | ||||
| <Operation> | Service code should be input in F06 prior to use this mode. | ||||
|
| Input F32 and press SET key to display ‘000’ (Initial setting). | ||||
|
| 000: Not notify |
|
|
|
|
|
| 001: Notify |
|
|
|
|
|
| Set your desired setting with | key and |
| key pressed. | |
|
| Press SET key to store the value and revert to PV display. | ||||
F33: | <Purpose> | Simply passing through if entered by mistake. | ||||
| <Operation> | Press SET key in “F33” displayed to revert to PV display. | ||||
|
| Buzzer tone is emitted if service code is not input in F06. | ||||
F34~F39 |
| Unused |
|
|
|
|
| <Operation> | Input F34~F39 and press SET key. |
|
|
8.Differential (The point of compressor ON and OFF )
COMP ON: | SV +0.4 |
COMP OFF: | SV |
9.Offset value
PV+1.0 is the offset value to adjust the difference between temperature of temp. sensor and the center of chamber.
20