7.Function mode
F00 Automatically revert to PV display
F01 | SVH (high temp. alarm SV) setting |
|
|
|
|
F02 | SVL (low temp. alarm SV) setting |
|
|
|
|
F03 | Battery accumulating time display |
|
|
|
|
F04 | Automatically revert to PV display |
|
|
|
|
F05 | Compressor delay time setting |
|
|
|
|
F06 | Service code input (code: 384) |
|
|
|
|
F07 * | Temperature Zero Adjustment |
|
|
|
|
F08 * | Automatically revert to PV display |
|
|
|
|
F09 * | Factory test mode …… Unavailable |
|
|
| |
F10 * | Factory test mode …… Unavailable |
|
|
| |
F11 * | Factory test mode …… Unavailable |
|
|
| |
F12 * | Display of temperature of temp. sensor |
|
|
| |
F13 * | Automatically revert to PV display |
|
|
|
|
F14 * | Display of temperature of Comp. sensor |
|
|
| |
F15 * | Automatically revert to PV display |
|
|
|
|
F16 * | Automatically revert to PV display |
|
|
|
|
F17 * | Model code setting (Initialization of | ||||
F18 * | Automatically revert to PV display |
|
|
|
|
F19 * | Automatically revert to PV display |
|
|
|
|
F20 * | Automatically revert to PV display |
|
|
|
|
F21 | Communication ID set |
|
|
|
|
F22 | Communication mode set |
|
|
|
|
F23 * | Automatically revert to PV display |
|
|
|
|
F24 | Remote alarm terminal output |
|
|
|
|
F25 |
|
|
|
| |
F26 * | “000” is displayed…Unavailable |
|
|
|
|
F27 * | “000” is displayed…Unavailable |
|
|
|
|
F28 * | Display of delay time of permission for measuring running rate (2 hrs timer) | ||||
F29 * | Display of delay time of permission for measuring running rate (8 hrs timer) | ||||
F30 * | ROM version is displayed |
|
|
|
|
F31 * | Automatically revert to PV display |
|
|
|
|
F32 * | Notice for condensing fan motor accumulating time |
|
| ||
F33 * | Automatically revert to PV display |
|
|
|
|
F34~F39 | Unused |
|
|
|
|
Setting | In PV display, keep pressing | key over 5seconds to display “F00”. | |||
procedure: | Input the desired function code with | key and |
| key. | |
| |||||
| Press SET key to be function mode available. |
|
|
*… You should input service code (384) in F06 beforehand. To cancel service code, input “000” in F06 or turn the power off.
F00: | <Purpose> | Simply passing through if entered by mistake. |
|
| <Operation> | Press SET key in “F00” displayed to revert to PV display. | |
F01: | <Purpose> | SVH (high temp. alarm SV) setting |
|
| <Operation> | Input F01 and press SET key to display “010” (initial value). | |
|
| Set the value in the range “005”~“020” with | key. Press SET key |
|
| to store the value and revert to PV display. |
|
F02: | <Purpose> | SVL (low temp. alarm SV) setting |
|
| <Operation> | Input F02 and press SET key to display | |
|
| Set the value in the range | key. Press SET key to |
|
| store the value and revert to PV display. |
|
F03: | <Purpose> | Battery accumulating time is displayed |
|
| <Operation> | Input F03 and press SET key to display alternately F03 with “00.0” | |
|
| (in case battery used for a month or less). Press SET key to revert to | |
|
| PV display. |
|
17