13.Fan motor operation and compressor protection

13.Fan motor operation and compressor protection

Fan motor operation :

1.When the temperature in Comp. sensor is lower than 20

;

 

Fan motor turns off with the power supplied

Turns on when the

 

temperature in Comp. sensor is higher than 80

 

Turns off when

 

the temperature in Comp. sensor is lower than 20

Turns on

 

when the temperature in Comp.sensor is higher than 80

 

 

2. When the temperature in Comp. sensor is lower than 20

;

 

Fan motor turns on with the power supplied

Turns off when the

 

temperature in Comp. sensor is lower than 20

 

Turns on when

 

the temperature in Comp. sensor is higher than 80

 

 

Compressor protection:

Compressor stops operation when the temperature in Comp. sensor

(during Fan Lock)

reaches 100 . (Compressor protection)

 

 

 

 

Starts operation again when the temperature in Comp. sensor is lower

 

than 65 .

 

 

 

15.Other specifications

(1)Lamp specifications: <Control PCB>

DP1: Orange lamp

 

Goes off: High/low temp. alarm, sensor failure, power failure

 

Lit

: Not in alert condition

DP2:

Green lamp

 

Goes off: Compressor does not activate. (normal condition)

 

Lit

: Compressor activates.

DP4:

Yellow lamp

 

Goes off: Fan motor does not activate. (normal condition)

<Display PCB>

Lit

: Fan motor activates.

 

 

DP51:

Red lamp

 

Goes off: Not in alarm condition (normal condition)

 

Blinks

: High/low temp. alarm (without delay), or sensor failure,

 

 

or power failure

DP53:

Orange lamp

 

Goes off: Battery accumulation time is less than 2.8yrs.

 

Lit

: Battery accumulation time is over than 2.8yrs (Expired).

 

Blinks

: In F11 performed

(3) Buzzer specifications:

 

 

High/low temp. alarm:

Intermittent tone emitted with delay

Sensor failure:

 

Intermittent tone emitted when any error code among

 

 

E01, E02, E05 and E06 is displayed

Power failure:

 

Intermittent tone emitted

Key operation:

 

A click emitted if available

Invalid input:

 

Alarming tone emitted for 1 second

22