Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2
9-20
Table 9-6: Test Channel Readings (Continued)
TEST Channel Minimum* Maximum* Description
RCELL TEMP 0C 70C Reaction Cell temperature is set to 50C. At the
setpoint, a typical reading is 3600 mV.
BLOCK TEMP 0C 70C The Block temperature is set to 50C. At the
setpoint, a typical reading is 3600 mV.
IZS TEMP 0C 70C The IZS temperature is set to 50C. At the
setpoint, a typical reading is 3600 mV.
CONV TEMP 0C 1000C The Converter temperature is 315C. At the
setpoint, a typical voltage is 3150 mV.
PMT TEMP 0C 70C The PMT temperature is unique in that the
voltage is inverse to the temperature. A typical
reading for 8C would be 4200 mV.
CHASSIS
TEMP 0C 70C The Chassis (Box) temperature is variable due
to variable ambient air temperature. The Box
temp generally runs about 5C above the
surrounding air temp. Thus in a 25C room, the
Box temp would be about 30C and have a
TEST channel voltage of about 2000 mV.
DCPS
VOLTAGE 0 mV 5000 mV The DCPS is a composite of several DC power
supply voltages in the instrument. It has been
arbitrarily set at 2500 mV, which is typical.
HVPS
VOLTAGE 0 V 5000 V The HVPS voltage is a scaled up reading of the
programming voltage going to the HVPS. Zero
to 1000mV corresponds 0-1000 VDC for the
HVPS, which is the maximum voltage possible.
A typical reading would be 700 mV
corresponding to 700 VDC for the HVPS.
*Minimum and Maximum readings depend on the DAC 3 switch settings of the V/F board. For the
standard 5VDC range, minimum corresponds to 0VDC and maximum corresponds to 5VDC.
9.1.6. Factory Calibration Procedure The Factory Cal procedure balances the PMT, preamp, and software gain factors so the
instrument has optimum noise, linearity, and dynamic range. It should be used in the cases where
you were unable to zero or span the instrument or slope and offset values were outside of the
acceptable range and other more obvious reasons for problems have been eliminated.