Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 5-32 Table 5-12: RS-232 Command Summary
Commands Definition
? Print help screen
T SET ALL Enable display of all test variables during T LIST
T SET name Display only NAME during T LIST
T LIST [ALL] Print all test variables enabled with T SET or ALL warnings
T name Print single test, "name" from Table 5-15
T CLEAR ALL Disable T LIST, use with T SET name
W SET ALL Enable display of all warnings during W LIST
W LIST [ALL] Print warnings enabled with W SET or ALL warnings
W name Print individual "name" warning from Table 5-16
W CLEAR ALL Disable W LIST, use with W SET
C LIST Print calibration commands
C command Execute calibration "command" from Table 5-18
D LIST Prints all I/O signal values
D name Prints single I/O signal value/state
D name=value Sets variable to new "value"
D LIST NAMES Lists diagnostic test names
D ENTER name Enters and starts 'name' diagnostic test
D EXIT Exits diagnostic mode
D SYS_RESET Resets analyzer(same as power-on)
D RAM_RESET System reset, plus erases RAM. Initializes DAS, NO, NOx,
NO2 conc readings, calib not affected.
D EE_RESET System reset, plus erases EEPROM (RAM_RESET actions +
setup variables, calibration to default values.
V LIST Print all variable names
V name Print individual "name" variable
V name=value Sets variable to new "value"
V CONFIG Print analyzer configuration
R count Print last "count" DAS averages