Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2
2-2
WARNING
Analyzer Exhaust – O3 Scrubber – Pump Pack
Danger – Analyzer exhaust contains ozone.
Do not defeat the internal zone scrubber. This device must always
be present between the analyzer reaction cell and pump.
Vent pump exhaust to well ventilated area at atmospheric pressure.
FIRE or EXPLOSION HAZARD
If the optional charcoal NO2 scrubber is used on the exhaust,
charcoal treated with halogen compounds must not be used.
Use only Teledyne API P/N 00596 charcoal.
Wait at least 5 minutes after turning off pump before
removing NO2 scrubber.
WARNING
Lethal voltages present inside case.
Do not operate with cover off during normal operation.
Before operation check for correct
input voltage and frequency.
Do not operate without proper chassis grounding.
Do not defeat the ground wire on power plug.
Turn off analyzer power before disconnecting
electrical subassemblies.