Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2
9-40
10 V output or non-isolated current loop
Switch State Comment
3 ON Use for non-isolated current loop or 10 V output
4 OFF Use for non-isolated current loop or 10 V output
5 OFF Use for non-isolated current loop or 10 V output
6 OFF Use for non-isolated current loop or 10 V output
5 V output or isolated current loop
Switch State Comment
3 OFF Use for isolated current loop or 5 V output
4 ON Use for isolated current loop or 5 V output
5 OFF Use for isolated current loop or 5 V output
6 OFF Use for isolated current loop or 5 V output
1 V output
Switch State Comment
3 OFF Use for 1 V output
4 OFF Use for 1 V output
5 ON Use for 1 V output
6 OFF Use for 1 V output
100 mV output
Switch State Comment
3 OFF Use for 100 mV output
4 OFF Use for 100 mV output
5 OFF Use for 100 mV output
6 ON Use for 100 mV output
1. After the switches and jumpers are set, turn on instrument power and complete the following:
A. Press SETUP-MORE-DIAG, then press ENTR. Scroll to D/A CALIBRATION, press
ENTR. Press A/D to select the first task, which is to calibrate the A/D converter.
B. Connect a DVM ground lead to TP3-AGND on the V/F board. Connect the positive lead
to TP9-DAC0.