Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2 2-10 Zero/Span Calibration Procedure (Continued)
Step For Units w/o
O2 Option For Units with
O2 Option Comment
13. Wait 10 min Wait for reading to stabilize at the zero value.
If you wait less than 10 minutes the final zero
value may drift.
14. Press ZERO The ZERO button will be displayed.
15. Press ENTR Pressing ENTR actually changes the
equations and zeros the O2 channel of the
instrument.
16. Press EXIT and
input O2 Span gas M200AH returns to the CAL menu. Now
switch gas streams to span gas.
17. Wait 10 min Wait for reading to stabilize at the span
value.
18. Press SPAN The SPAN button should be displayed. If
there is no SPAN button check the
Troubleshooting Section 9.2.8 for
instructions on how to proceed. In certain
circumstances at low span gas concentrations
both the ZERO and SPAN buttons will
appear.
19. Press ENTR Pressing ENTR actually changes the
equations so that the concentration displayed
is the same as the expected span
concentration you entered above, thus
spanning the instrument.
20. Press EXIT Pressing EXIT returns the instrument to
SAMPLE mode.