Step For Units w/o
O2 Option For Units with
O2 Option Comment
1. Press
CAL-CONC-NOX Press
CAL-NOX-ENTR
CONC-NOX
This key sequence causes the M200AH to
prompt for the expected span concentration.
Enter the span value by pressing the key under
each digit until the expected value is set.
2. Press ENTR Press ENTR ENTR stores the expected NOx span value. The
internal formulas are adjusted to compute this
number when span gas concentration is input
into the instrument.
3. Press
CAL-CONC-NO Press
CAL-NOX-ENTR
CONC-NO
In the same CAL-CONC sub menu press the
NO button and enter the expected NO span
value. As before this value will be used in the
internal formulas to compute the NO
concentration value.
4. Press ENTR Press ENTR ENTR stores the expected NO span value. The
internal formulas are adjusted to compute this
number when span gas concentration is input
into the instrument.
5. Press
CAL-O2-CONC Enter the expected O2 span value by pressing
the key under each digit until the expected
value is set.
6. Press ENTR ENTR stores the expected O2 span value. The
internal formulas are adjusted to compute this
number when span gas concentration is input
into the instrument.
7. Press EXIT Press EXIT Returns instrument to SAMPLE mode.
8. Press
SETUP-RNGE-
MODE-SNGL
Press
SETUP-RNGE-
MODE-SNGL
If necessary, you may want to change ranges.
Normally the instrument is shipped in single
range mode set at 100 ppm for NOx and 25%
for O2. We recommend doing the initial
checkout with these ranges.
9. Press
SETUP-RNGE-SET Press
SETUP-RNGE-SET After SETUP-RNGE-SET, enter 100 and press
ENTR. The instrument will now be in the
100 ppm range for NOx and (if installed) will
display the O2 range. Enter 25.0 and press
ENTR.