Teledyne API Model 200AH NOX Analyzer Instruction Manual, 06492, Rev. G2

3-1

3. SPECIFICATIONS, WARRANTY

3.1. Specifications

NOX Specifications O2 Option
Operating Modes NO/NOx switching mode, NO only
mode, NOx only mode. Paramagnetic Sensor
Ranges In 1 ppm increments from 5 ppm to
5000 ppm; Single range, independent
ranges or autoranging
In 1 % increments from 5 % to
100%, Independent of NOX range
Noise at zero <0.020 ppm RMS 0.05% O2
Noise at span <0.5% of reading RMS above 20 ppm <0.1% O2
Detection Limit (Note 1) 0.040 ppm 0.1% O2
Zero Drift (Note 2) <0.5% full scale/24 hours 0.1% O2
Zero Drift (Note 2) <1.0% full scale/7 days 0.1% O2
Span Drift (Note 2) <1% FS/24 hours 0.1% O2
Lag Time
- Switching Mode - 20 sec (Note 3)
- NOx mode - 4 sec (Note 3)
< 4sec
Response Time
- Switching Mode - 95% in < 40 sec (Note 3)
- NOx mode - 95% in < 10 sec (Note 3)
95% in <10 sec
Sample Flow Rate (Analyzer) 370 cc/min 10% (Including bypass)
Sample Flow Rate (Sensors) 40 cc/min 10% 80 cc/min 10%
Linearity 1% of full scale 0.1% O2
Precision 0.5% of reading 0.1% O2
Temperature Range 5-40C
Humidity 0-95% RH non-condensing
Temp Coefficient < 0.1% per C < 0.1% per C
Voltage Coefficient < 0.1% per V < 0.1% per V
Dimensions HxWxD 7"x17"x23.6" (18 cm x 43 cm x 61 cm)
Weight, Analyzer 45 lbs (22 kg)
Weight, Pump Pack 16 lbs (7 kg)
Power, Analyzer 100 V~ 50/60 Hz, 120 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 200 watts
Power, Analyzer4 230 V~ 50 Hz, 2.5A
Power, Ext Pump 110 V~ 60 Hz, 220 V~ 50 Hz, 240 V~ 50 Hz, 295 watts
Power, Ext Pump4 230 V~ 50 Hz, 2.5A
Environmental Installation Category (Over-voltage Category) II
Pollution Degree 2
Analog Resolution 1 part in 2048 of selected voltage or
current range 1 part in 2048 of selected voltage or
current range
Recorder Output 0-100mV, 0-1, 5, 10v, bipolar 0-100mV, 0-1, 5, 10v, bipolar
Current Loop Option 4-20ma isolated 4-20ma isolated
Status 12 Status Outputs from opto-isolator
Measurement Units ppm, mg/m3 %
1. Defined as twice the zero noise level.
2. At constant temperature and voltage.
3. Lag & response times longer for external converter option.
4. Electrical ratings for CE Mark compliance.