2101 Field Wizard Section 5 Maintenance

2101 Field Wizard Section 5 Maintenance

5.3 Desiccant

The Field Wizard uses desiccant to protect the internal compo- nents from moisture damage. The cartridge is filled with indi- cating silica gel that is blue or yellow when dry. As the desiccant becomes saturated, the color changes from blue to pink, or from yellow to green. Replace the desiccant before the entire length of the cartridge turns pink or green.

5.3.1 Replacing the

The desiccant is contained in a cartridge located on the left side

Desiccant: 2101

of the Field Wizard. To remove the cartridge, unscrew the collar

 

and slide the cartridge out of the Field Wizard. The clear tube

 

reveals the silica gel desiccant inside.

Collar

To replace the silica gel desiccant:

 

1.

Hold the cartridge upright with the collar at the top.

 

2.

As shown in the margin, push the collar off the cartridge.

 

3.

Empty the saturated silica gel beads or granules.

 

4.

Fill the tube with new (Isco P/N 099-0011-03) or reacti-

 

 

vated (see section 4.3.2) silica gel desiccant.

 

5.

Press the collar onto the tube.

 

6.

Slide the cartridge into the Field Wizard. Tighten the col-

 

 

lar to seal the cartridge in place.

Cartridge

 

 

5.3.2 Reactivating the

Silica gel beads and granules of desiccant can be reactivated.

Desiccant

 

CAUTION

 

 

 

Desiccant may produce irritating fumes when heated. Observe

 

the following precautions:

 

Use a vented oven in a well-ventilated room.

 

Do not remain in the room while the regeneration is taking

 

 

place.

 

Use the recommended temperature. Avoid heating the

 

 

desiccant at higher than recommended temperatures.

 

There have been reports of irritating fumes coming from the des-

 

iccant during reactivation. While our attempts to duplicate the

 

problem have been unsuccessful, we still urge you to use caution.

 

Material Safety Data Sheets are in the back of this manual.

 

The desiccant’s ability to remove moisture may lessen with each

 

saturation/reactivation cycle, resulting in a need for more fre-

 

quent service. After several cycles, the desiccant may no longer

 

be effective as it saturates too quickly. At this point, replace the

 

desiccant.

Silica gel

To reactivate the silica gel desiccant, pour the spent desiccant

 

into a heat resistant container. Never heat the cartridge

 

assembly; it will melt. Heat the silica gel in a vented convection

5-2