7.3.TRANSDUCER CONNECTIONS

Connectors J1, 2, 3 and 4 (Item 7) are 15 pin D style connectors wired in the standard Hastings Instruments pin-out (H pin-out).

J1, J2, J3, J4

1

NC

 

 

2

NC

Valve Cntrl

 

 

 

Voltage

 

3

NC

mA Sig

 

4

NC

mA Sig

 

5

Sig. Com.

 

Transducer

6

Sig. In

 

7

Case Gnd.

 

Connectors

 

8

Cntrl Over-ride

1.5mA

(DB-15)

9

-15 VDC

 

(H-Pinout)

 

10

NC

 

 

 

 

11

+15 VDC

 

 

12

Valve Return

 

 

13

NC

Ext-In

 

14

Set Point Out

 

 

15

+5 VDC Ref.

Not Used.

7.4.ALARMS

Connector J8 (item 9) provides the user with open-collector, opto-isolated alarms for individual channels. Each channel is provided with one user settable “High” and one “Low” alarm.

J8

1

Chnl 1, High Alarm

 

2

Chnl 1, Low Alarm

 

3

Chnl 1, Alarm Return

 

4

Chnl 2, High Alarm

 

5

Chnl 2, Low Alarm

 

6

Chnl 2, Alarm Return

 

7

NC

 

8

NC

 

9

NC

 

10

NC

Open

11

NC

Collector,

12

NC

Opto-

13

NC

Isolated, High

14

NC

& Low

15

NC

Alarms

16

NC

(HD DB-26)

17

NC

 

18

NC

 

19

Chnl 3, High Alarm

 

20

Chnl 3, Low Alarm

 

21

Chnl 3, Alarm Return

 

22

Chnl 4, High Alarm

 

23

Chnl 4, Low Alarm

 

24

Chnl 4, Alarm Return

 

25

NC

 

26

NC

164-122012 PowerPOD 400