Congratulations!

You have just calibrated your PowerPod-400to read zero at zero volts and the transducer’s max display value at the transducer’s max signal input.

The only thing left to do is to insure that your transducer is calibrated and that it is wired correctly to the PowerPod-400.

*Individual channels may be calibrated with calibrated transducers connected, breakout connectors (not supplied) and precision multi-meters with NIST traceable calibrations (also not supplied). Such procedures are not covered in this document.

!If the calibration for the channel just zeroed had been severely compromised prior to beginning this procedure, zero may not be immediately displayed and you may have to repeat the procedure after resetting the span. Check that the span has been set correctly!

#If a signal level within .1% of the maximum DC voltage level is not attainable, the power supply may need to be repaired.

164-122012 PowerPOD 400

Page 40 of 42