3.After insert liner between the see through type gap sensor, press the

button under “V” icon to continue.

4.Press the button under “+” icon to increase Stock AD value by

50 increment. And press the button under “V” icon to calibrate the sensor sensitivity.

GAP Calibrate 185

+ - V

4.1.2.1.2 Manual Gap:

If Auto Gap sensor calibration can not apply to the media, please use Manual Gap function to calibrate sensor manually. This function can be used to fine tune sensor sensitivity. When enter [Manual Gap] option, you will see following message.

Gap Sensor

Stock AD = 50

Intension = 140

+ - V

Stock AD on liner

Gap Sensor

Stock AD = 163

Intension = 140

+ - V

Stock AD on paper

Press the button under “+” and “-“ to increase and decrease

sensor intension, and press the under “V” to set the value to printer.

When adjust the sensor sensitivity manually, we can get the different Stock AD on liner and paper in each sensor intension. The difference between these two values must be larger than 50 levels then printer can register label position correctly.