Users Manual 63
PART 3 TC218 2. Specifications
2.2 Internal block diagram
The TC218 performs the following operations.The external analog signals come to the buffer amplifier through the filter. The multiplexersequentially selects CH1 to CH8 to convert the input analog signals into digital data via the A/Dconverter. The converted digi tal data reaches to the internal control circuit through opticalisolator. Every time when the T2 CPU requests to read the converted data, the internal controlcircuit sends the data to the T2 CPU. The TC218's parameters are stored in the EEPROM.
Filter
Buffer
1
3
2
Jumper
Filter
Buffer
1
3
2
Jumper
A/D conversion
Optical isolation
CH1 P
N
Internal control circuit
CH8 P
N
Filter
DC/DC
converter
Voltage
check
circuit
Regulator +5V
+15V
-15V
AG
P24
COM
LG
FG
EEPROM
LED
T2 CPU
Reference vol tage
+15V
Filter
Buffer
1
3
2
Jumper
CH2 P
N
+15V
Multiplexer
CTi Automation - Phone: 800.894.0412 - Fax: 208.368.0415 - Web: www.ctiautomation.net - Email: info@ctiautomation.net