Installation and Configuration

Remote/Local Operation

Table 2.10 Remote Mode Power On Conditions

Condition

Default Settings

7.5-140 Model Example

 

 

 

Voltage

0 V

VSET 0

 

 

 

Current

0 A

ISET 0

 

 

 

Soft Voltage Limit

VMAX (see models)

VMAX 7.5

 

 

 

Soft Current Limit

IMAX (see models)

IMAX 140

 

 

 

OVP Trip Voltage

Model VMAX + 10%

OVSET 8.25

 

 

 

Delay

0.5 s

DLY 0.5S

 

 

 

Remote Enable

ON

REN ON

 

 

 

Foldback Protection

OFF

FOLD OFF

 

 

 

Output

ON

OUT ON

 

 

 

Hold

OFF

HOLD OFF

 

 

 

Unmask

NONE

UNMASK NONE

 

 

 

AUXA

OFF

AUXA OFF

 

 

 

AUXB

OFF

AUXB OFF

 

 

 

Remote Mode While in remote mode, use the interface commands to control the output of the Operation power supply from a computer. See Section 3 for a complete list of

device-dependent commands available with this interface.

From remote mode, change to local mode operation by pressing the front panel LOCAL button or by sending either the GTL (Go to Local) command or the REN (Remote Enable) OFF command. You can disable the LOCAL button by using the LLO (Local Lockout) command. For an example of how to use Local Lockout, see “Local Lockout (LLO) Command” on page 30.

To return to remote mode, ensure that the REN command is set to ON and then send any valid device-dependent command. Since the remote mode output settings may be different from the local mode settings, the output is programmed to protect the load by turning off when you toggle from local to remote mode during a session. Check your remote settings and then send OUT ON to restore the output.

28

Operating Manual for RS-232 for XHR/XFR Series Power Supply

Page 30
Image 30