A-18 975-0200-01-01
SCPI Command Reference
Table A-4 SCPI Commands for Calibration
Function SCPI Command Description Parame ter an d
Range Query
Restore
Factory
Calibration
[:]CALibration[<channel>]:RESTore Restores the calibration to
the factory defaults. N/A N/A
Calibrate
OVP
protection.
[:]CALibration[<channel>][:VOLTage]
:PROTection[:OVER] Calibrate the over voltage
protection. <OVP Voltage Set
in Volts> No
Enter Output
Voltage Data [:]CALibration[<channel>]:OUTPut
:VOLTage Calibrates the output
voltage. <Voltage set in
volts> No
Enter Output
Current Data [:]CALibration[<channel>]:OUTPut
:CURRent Set voltage output
calibration level <Current set in
amps> No
Calibrate the
Voltage
controlled
Non Isolated
Voltage APG
feature
[:]CALibration[<channel>]:MONitor
[:VOLTage] Calibrate the voltage
monitor signal system. ? | <NR1> |
<nothing for
interactive mode>
Yes
Calibrate the
Voltage
controlled
Isolated
Voltage APG
feature
[:]CALibration[<channel>]:MONitor
[:VOLTage]:ISOLated Calibrate the isolated
voltage monitor signal
system.
? | <NR1> |
<nothing for
interactive mode>
Yes
Calibrate the
Voltage
controlled
Non Isolated
Current APG
feature
[:]CALibration[<channel>]:MONitor
:CURRent Calibrate the current
monitor signal system. ? | <NR1> |
<nothing for
interactive mode>
Yes