Analog Programming (APG) and Isolated Analog Programming (ISOL)
4-12 975-0200-01-01
Voltage-Controlled Current APG Setup

Activating APG Current Mode

To activate APG current mode using an external voltage source:
1. Turn the mode control knob to the CAP position or press the Rotary
knob/Enter button if the control knob is already at the CAP position.
CAPr is displayed on the output voltage display.
2. Turn the Rotary knob/Enter button to select source Un1S (Voltage
source from non-isolated connector).
3. Press the Rotary knob/Enter button to commit the selected mode.
LE C is displayed on the output voltage display.
4. Select the input source level using the Rotary knob/Enter button.
5. Once the desired level has been selected, press the Rotary knob/Enter
button to commit the change.
The display will return to the normal display.

Deactivating APG Current Mode

To deactivate the APG current mode:
1. Turn the mode control knob to the CAP position or press the Rotary
knob/Enter button if the control knob is already at the CAP position.
2. CAPr is displayed on the output voltage display.
3. Turn the Rotary knob/Enter button to select nonE.
4. Press the Rotary knob/Enter button to commit the change.
The display will return to the normal display.

Query for Analog Current Input Level

To query for analog current input level from non-isolated input:
1. Turn the mode control knob to the CAP position or press the Rotary
knob/Enter button if the control knob is already at the CAP position.
CAPr is displayed on the output voltage display.
2. If the mode is currently nonE then select any other APG mode. If this
step is done, be sure to restore the APG setting to nonE when the
operation is complete. Otherwise, do not rotate the Rotary knob/Enter
button to keep the current APG mode.