Emerson 644 manual Table A-2. Ambient Temperature Effect, Thermocouples

Page 53

Reference Manual

00809-0400-4728, Rev AA June 2011

Rosemount 644

Ambient Temperature Effect

Table A-2. Ambient Temperature Effect

 

 

 

Temperature Effects per 1.0 °C

 

 

 

 

 

Input Range

(1.8 °F) Change in Ambient

 

 

 

Sensor Options

Sensor Reference

(°C)

Temperature(1)

Range

D/A Effect(2)

2-, 3-, 4-wire RTDs

 

 

 

 

 

 

 

Pt 100 (α = 0.00385)

IEC 751

-200 to 850

0.003

°C (0.0054 °F)

Entire Sensor Input Range

0.001%

of span

Pt 200 (α = 0.00385)

IEC 751

-200 to 850

0.004

°C (0.0072 °F)

Entire Sensor Input Range

0.001% of span

Pt 500 (α = 0.00385)

IEC 751

-200 to 850

0.003

°C (0.0054 °F)

Entire Sensor Input Range

0.001% of span

Pt 1000 (α = 0.00385)

IEC 751

-200 to 300

0.003

°C (0.0054 °F)

Entire Sensor Input Range

0.001% of span

Pt 100 (α = 0.003916)

JIS 1604

-200 to 645

0.003

°C (0.0054 °F)

Entire Sensor Input Range

0.001% of span

Pt 200 (α = 0.003916)

JIS 1604

-200 to 645

0.004

°C (0.0072 °F)

Entire Sensor Input Range

0.001% of span

Ni 120

Edison Curve No. 7

-70 to 300

0.003

°C (0.0054 °F)

Entire Sensor Input Range

0.001% of span

Cu 10

Edison Copper

-50 to 250

0.03 °C (0.054 °F)

Entire Sensor Input Range

0.001% of span

 

Winding No. 15

 

 

 

 

 

 

Pt 50 (α = 0.00391)

GOST 6651-94

-200 to 550

0.004

°C (0.0072 °F)

Entire Sensor Input Range

0.001% of span

Pt 100 (α = 0.00391)

GOST 6651-94

-200 to 550

0.003

°C (0.0054 °F)

Entire Sensor Input Range

0.001% of span

Cu 50 (α = 0.00426)

GOST 6651-94

-50 to 200

0.008

°C (0.0144 °F)

Entire Sensor Input Range

0.001% of span

Cu 50 (α = 0.00428)

GOST 6651-94

-185 to 200

0.008

°C (0.0144 °F)

Entire Sensor Input Range

0.001% of span

Cu 100 (α = 0.00426)

GOST 6651-94

-50 to 200

0.004

°C (0.0072 °F)

Entire Sensor Input Range

0.001% of span

Cu 100 (α = 0.00428)

GOST 6651-94

-185 to 200

0.004

°C (0.0072 °F)

Entire Sensor Input Range

0.001% of span

Thermocouples

 

 

 

 

 

 

 

 

 

 

0.014

°C

T 1000 °C

0.001%

of span

Type B

NIST Monograph 175,

100 to 1820

0.032

°C – (0.0025% of (T –

300 °C T < 1000 °C

0.001% of span

IEC 584

300))

 

 

 

 

 

 

 

 

 

 

 

 

 

0.054

°C – (0.011% of (T – 100))

100 °C T < 300 °C

0.001% of span

Type E

NIST Monograph 175,

-50 to 1000

0.005

°C + (0.0043% of T)

All

0.001% of span

IEC 584

 

 

 

 

 

 

 

 

 

 

 

 

 

NIST Monograph 175,

 

0.0054 °C + (0.00029%of T)

T0 °C

0.001% of span

Type J

-180 to 760

0.0054 °C + (0.0025% of

T < 0 °C

0.001% of span

IEC 584

 

 

absolute value T)

 

 

 

 

 

 

 

 

 

 

NIST Monograph 175,

 

0.0061 °C + (0.0054% of T)

T 0 °C

0.001% of span

Type K

-180 to 1372

0.0061 °C + (0.0025% of

T < 0 °C

0.001% of span

IEC 584

 

 

absolute value T)

 

 

 

 

 

 

 

 

 

Type N

NIST Monograph 175,

-200 to 1300

0.0068 °C + (0.00036% of T)

All

0.001% of span

IEC 584

 

 

 

 

 

 

 

 

 

 

 

 

Type R

NIST Monograph 175,

0 to 1768

0.016

°C

T 200 °C

0.001% of span

IEC 584

0.023

°C – (0.0036% of T)

T < 200 °C

0.001% of span

 

 

Type S

NIST Monograph 175,

0 to 1768

0.016

°C

T 200 °C

0.001% of span

IEC 584

0.023

°C – (0.0036% of T)

T < 200 °C

0.001% of span

 

 

 

NIST Monograph 175,

 

0.0064 °C

T0 °C

0.001% of span

Type T

-200 to 400

0.0064 °C +(0.0043% of

T < 0 °C

0.001% of span

IEC 584

 

 

absolute value T)

 

 

 

 

 

 

 

 

 

 

 

 

0.0054 °C + (0.00029% of T)

T0 °C

0.001% of span

DIN Type L

DIN 43710

-200 to 900

0.0054 °C + (0.0025% of

T < 0 °C

0.001% of span

 

 

 

absolute value T)

 

 

 

 

 

 

0.0064 °C

T0 °C

0.001% of span

DIN Type U

DIN 43710

-200 to 900

0.0064 °C + (0.0043% of

T < 0 °C

0.001% of span

 

 

 

absolute value T)

 

 

 

Type W5Re/W26Re

ASTM E 988-96

0 to 2000

0.016

°C

T 200 °C

0.001% of span

0.023

°C – (0.0036% of T)

T < 200 °C

0.001% of span

 

 

 

 

 

 

0.007

°C

T 0 °C

0.001% of span

GOST Type L

GOST R 8.585-2001

-200 to 800

0.007

°C – (0.003% of absolute

T < 0 °C

0.001% of span

 

 

 

value T)

 

 

 

A-7

Image 53
Contents 00809-0400-4728, Rev AA June Reference ManualPage Table of Contents Appendix a Specifications Reference DataAppendix B TOC-2Customer Central National Response Center International Rosemount Temperature TransmittersPage Section Introduction Safety MessagesExplosions could result in death or serious injury Process leaks could result in death or serious injuryTransmitter OverviewManual Considerations 250 C 482 F Oven Temperature 5 6 7 8 Extension Length 22 20 815 C 1500 F Oven TemperatureExample AboveSection Installation Simulate Sensor Input Within Specifications? Refer to ConfigurationBench Calibration? Mounting From draining into the transmitter housingMounting a 644H to a DIN Rail Rail asymmetricHead Mount Transmitter with DIN Plate Style Sensor InstallationTypical European Installation Head Mount Transmitter with Threaded Sensor Typical North American InstallationInstalling the LCD Display LCD Display InstallationPower terminals does not drop below 9 Vdc WiringThermocouple or Millivolt Inputs Sensor ConnectionsRTD or Ohm Inputs Sensor Lead Wire Resistance Effect- RTD InputSensor Input Approximate Basic Error Error due to amb. temp. var. of ± 25 C Pt100 2-wire RTDFoundation fieldbus Installation Power SupplyGround the Transmitter Ungrounded Thermocouple, mV, and RTD/Ohm InputsGrounded Thermocouple Inputs Overview Safety Messages Section ConfigurationDevice Description Node Address General Block InformationModes AutoResource Transducer Analog Input Other Link Active Scheduler Block InstallationFoundation Fieldbus Function Blocks CapabilitiesResource Block FeaturesUnicode ReportsFailedalarms FeaturesselMaxnotify PlantWeb AlertsFailedmask MaintalarmsRecommendedaction Advisory AlarmsRecommended Actions for PlantWeb Alerts Sensor Transducer Block Configure the AI blockAnalog Input AI Function Block DampingXdscale and Outscale Pressure Channel Temperature ChannelLtype Transducer Block Configuration ExamplesFiltering Process AlarmsOutd AlarmtypeDisplayparamsel LCD Transducer BlockBLKTYPE#1 PARAMINDEX#1Overview Operation and MaintenanceUNITSTYPE#1 CUSTOMUNITS#1Troubleshooting guide Troubleshooting GuidesProblems with communications flowchart Calpointlo + Calminspan Recall Factory Trim DiagnosticsBlock Blockerr messages Sensor Transducer Block Xderr messagesCalibrationerr DescriptionStatus Manual ModeSimulation SimulatePower Up Troubleshooting AI blockSymptom Possible Causes Recommended Actions 00809-0400-4728, Rev AA June AI Blockerr Conditions8lists conditions reported in the Blockerr parameter Block Errors10. Resource Block Condition NameAuto Self Test Procedure for the LCDSymptom Possible Causes Recommended Action FunctionalInputs Appendix a Specifications Reference DataSpecifications Physical Performance Rosemount Conformance to Specifications Foundation Fieldbus SpecificationsFunction Blocks 2 σ -1 σAlarms Power SupplyBackup Link Active Scheduler LAS Software Upgrade in the FieldAccuracy Example AccuracyTable A-1. Rosemount 644 Input Options and Accuracy Thermocouples Ambient Temperature EffectTable A-2. Ambient Temperature Effect Temperature Effects Examples 644H DIN a Head Mount Dimensional DrawingsMounting LCD Display00809-0400-4728, Rev AA JuneRing 76.2 27.9 74.4 25.4 44.5 70.0 Sanitary Housing CoversLCD Display Cover Options Ordering InformationAlarm Level Configuration Standard Configuration StandardAssemble To Options Standard Enclosure Options StandardConsiderations TaggingWith DeltaV, with AMS, or other Foundation fieldbus host or Transmitter ConfigurationConfiguration tool Custom Configuration Standard Foundation fieldbus ConfigurationRequirements Option Code Specification Standard Block ConfigurationAtex Directive 94/9/EC CE Electromagnetic Compatibility Compliance TestingAppendix B Product Certifications Rosemount 644 with Foundation fieldbus Hazardous Locations CertificatesNorth American Certifications Fisco Loop/Power Terminals Loop/Power TerminalsEuropean Certifications Special Conditions for Safe UseIECEx Certifications Temperature ClassTo 60 C Fisco Group IIC To 80 CTransmitter I.S Transmitter Fisco Sensor Table B-3. Entity ParametersTransmitter Sensor Schedule of LimitationsPattern Approval for Measuring Instruments See Certificate Ukraine GostApproval Temp Description Group CodeInstallation Drawings Rosemount Rosemount Rosemount 00809-0100-4728, Rev AA June Sheet 2 00809-0100-4728, Rev AA June Sheet 3 00809-0100-4728, Rev AA June 00809-0100-4728, Rev AA June Sheet 2 Rosemount Rosemount 00809-0100-4728, Rev AA June Sheet 2 00809-0100-4728, Rev AA June Sheet 3 Definition Basic Setup Resource BlockAppendix C Foundation fieldbus Block Information Index Parameters and DescriptionsFeatures FailedenableFeaturesel Maintenable00809-0400-4728, Rev AA Description Effects transmitter operation Sensor Transducer BlockParameters Descriptions Primaryvaluetype PrimaryvaluerangeSensrdetailedstatus CalvandusencoeffSESNOR1DAMPING XderrorTransducerdirectory TransducertypeAnalog Input AI Function Block Analog Measurement AccessOutd Cutoff Filter Convert CalcAckoption AI Parameter TableLolopri Parameter Index Description LCD Transducer BlockTable C-4. LCD Transducer Block Parameters and Descriptions Parameter Index PID BlockTable C-5. PID Block Parameters and Descriptions Index Bad Uncertain Ungrounded Wiring MODEBLK.TARGETLolim Lololim Lolopri Lopri Lower Trim Method DIN Rail Threaded SensorGrounded Ungrounded Wiring Limited Man Mode Unicode UNITSTYPE# Upper Trim MethodSecurity Self Test Direct Zero TrimIndex-4 Page Emerson Process Management Rosemount Inc Temperature GmbH Pacific Private Limited