Emerson 644 manual Temperature Effects Examples

Page 54

Rosemount 644

Reference Manual

00809-0400-4728, Rev AA June 2011

Table A-2. Ambient Temperature Effect

 

 

 

Temperature Effects per 1.0 °C

 

 

 

 

 

Input Range

(1.8 °F) Change in Ambient

 

 

 

Sensor Options

Sensor Reference

(°C)

Temperature(1)

Range

D/A Effect(2)

Other Input Types

 

 

 

 

 

 

Millivolt Input

 

-10 to 100 mV

0.0005 mV

Entire Sensor Input Range

0.001%

of span

2-, 3-, 4-wire Ohm

 

0 to 2000 Ω

0.0084 Ω

Entire Sensor Input Range

0.001% of span

(1)Change in ambient is with reference to the calibration temperature of the transmitter 68 °F (20 °C) from factory.

(2)Does not apply to FOUNDATION fieldbus.

Transmitters can be installed in locations where the ambient temperature is between –40 and 85 °C (–40 and 185 °F). In order to maintain excellent accuracy performance, each transmitter is individually characterized over this ambient temperature range at the factory.

Temperature Effects Examples

When using a Pt 100 (α = 0.00385) sensor input at 30 °C span at 30 °C ambient temperature:

Digital Temperature Effects: 0.003 °C x (30 - 20) = 0.03 °C

D/A Effects: No D/A effects apply

Worst Case Error: Digital + Digital Temperature Effects = 0.15 °C + 0.03 °C = 0.18 °C

• Total Probable Error:

0.152 + 0.032 = 0.153°C

A-8

Image 54
Contents Reference Manual 00809-0400-4728, Rev AA JunePage Table of Contents Appendix B Specifications Reference DataAppendix a TOC-2Rosemount Temperature Transmitters Customer Central National Response Center InternationalPage Explosions could result in death or serious injury Safety MessagesSection Introduction Process leaks could result in death or serious injuryOverview ManualTransmitter Considerations Example 22 20 815 C 1500 F Oven Temperature250 C 482 F Oven Temperature 5 6 7 8 Extension Length AboveSection Installation Within Specifications? Refer to Configuration Bench Calibration?Simulate Sensor Input Mounting a 644H to a DIN Rail From draining into the transmitter housingMounting Rail asymmetricInstallation Typical European InstallationHead Mount Transmitter with DIN Plate Style Sensor Typical North American Installation Head Mount Transmitter with Threaded SensorLCD Display Installation Installing the LCD DisplayWiring Power terminals does not drop below 9 VdcSensor Connections Thermocouple or Millivolt InputsSensor Lead Wire Resistance Effect- RTD Input Sensor Input Approximate Basic ErrorRTD or Ohm Inputs Pt100 2-wire RTD Error due to amb. temp. var. of ± 25 CGround the Transmitter Power SupplyFoundation fieldbus Installation Ungrounded Thermocouple, mV, and RTD/Ohm InputsGrounded Thermocouple Inputs Section Configuration Overview Safety MessagesModes General Block InformationDevice Description Node Address AutoLink Active Scheduler Block Installation Resource Transducer Analog Input OtherCapabilities Foundation Fieldbus Function BlocksUnicode FeaturesResource Block ReportsMaxnotify FeaturesselFailedalarms PlantWeb AlertsMaintalarms FailedmaskAdvisory Alarms Recommended Actions for PlantWeb AlertsRecommendedaction Analog Input AI Function Block Configure the AI blockSensor Transducer Block DampingPressure Channel Temperature Channel LtypeXdscale and Outscale Configuration Examples Transducer BlockProcess Alarms FilteringAlarmtype OutdBLKTYPE#1 LCD Transducer BlockDisplayparamsel PARAMINDEX#1UNITSTYPE#1 Operation and MaintenanceOverview CUSTOMUNITS#1Troubleshooting Guides Troubleshooting guideProblems with communications flowchart Calpointlo + Calminspan Block Blockerr messages DiagnosticsRecall Factory Trim Sensor Transducer Block Xderr messagesDescription CalibrationerrSimulation Manual ModeStatus SimulateSymptom Possible Causes Recommended Actions Troubleshooting AI blockPower Up 00809-0400-4728, Rev AA June AI Blockerr Conditions10. Resource Block Block Errors8lists conditions reported in the Blockerr parameter Condition NameSelf Test Procedure for the LCD Symptom Possible Causes Recommended ActionAuto Appendix a Specifications Reference Data SpecificationsFunctionalInputs Physical Performance Function Blocks Foundation Fieldbus SpecificationsRosemount Conformance to Specifications 2 σ -1 σBackup Link Active Scheduler LAS Power SupplyAlarms Software Upgrade in the FieldAccuracy Table A-1. Rosemount 644 Input Options and AccuracyAccuracy Example Ambient Temperature Effect Table A-2. Ambient Temperature EffectThermocouples Temperature Effects Examples Dimensional Drawings 644H DIN a Head Mount00809-0400-4728, Rev AA LCD DisplayMounting JuneSanitary Housing Covers LCD Display CoverRing 76.2 27.9 74.4 25.4 44.5 70.0 Ordering Information OptionsAssemble To Options Standard Configuration StandardAlarm Level Configuration Standard Enclosure Options StandardTagging ConsiderationsTransmitter Configuration Configuration toolWith DeltaV, with AMS, or other Foundation fieldbus host or Requirements Option Code Specification Standard Foundation fieldbus ConfigurationCustom Configuration Standard Block ConfigurationCE Electromagnetic Compatibility Compliance Testing Appendix B Product CertificationsAtex Directive 94/9/EC Hazardous Locations Certificates North American CertificationsRosemount 644 with Foundation fieldbus European Certifications Loop/Power TerminalsFisco Loop/Power Terminals Special Conditions for Safe UseTo 60 C Temperature ClassIECEx Certifications Fisco Group IIC To 80 CTransmitter Sensor Table B-3. Entity ParametersTransmitter I.S Transmitter Fisco Sensor Schedule of LimitationsApproval Temp Ukraine GostPattern Approval for Measuring Instruments See Certificate Description Group CodeInstallation Drawings Rosemount Rosemount Rosemount 00809-0100-4728, Rev AA June Sheet 2 00809-0100-4728, Rev AA June Sheet 3 00809-0100-4728, Rev AA June 00809-0100-4728, Rev AA June Sheet 2 Rosemount Rosemount 00809-0100-4728, Rev AA June Sheet 2 00809-0100-4728, Rev AA June Sheet 3 Basic Setup Resource Block Appendix C Foundation fieldbus Block InformationDefinition Parameters and Descriptions IndexFeaturesel FailedenableFeatures Maintenable00809-0400-4728, Rev AA Sensor Transducer Block Parameters DescriptionsDescription Effects transmitter operation Sensrdetailedstatus PrimaryvaluerangePrimaryvaluetype CalvandusencoeffTransducerdirectory XderrorSESNOR1DAMPING TransducertypeOutd Cutoff Filter Analog Measurement AccessAnalog Input AI Function Block Convert CalcAI Parameter Table AckoptionLolopri LCD Transducer Block Table C-4. LCD Transducer Block Parameters and DescriptionsParameter Index Description PID Block Table C-5. PID Block Parameters and DescriptionsParameter Index Index Lolim Lololim Lolopri Lopri Lower Trim Method Ungrounded Wiring MODEBLK.TARGETBad Uncertain DIN Rail Threaded SensorSecurity Self Test Limited Man Mode Unicode UNITSTYPE# Upper Trim MethodGrounded Ungrounded Wiring Direct Zero TrimIndex-4 Page Temperature GmbH Pacific Private Limited Emerson Process Management Rosemount Inc