Xantrex Technology ENET-XFR3 manual Remote Mode Operation

Page 43

Remote/Local Operation

Table 2-10Remote Mode Power On Conditions

VoltageCondition

Default0 V

Settings7VSET.5-140ModelExample

Current

0 A

 

ISET

 

 

 

Soft Voltage Limit

VMAX (see models)

VMAX 7.5

 

 

 

Soft Current Limit

IMAX (see models)

IMAX 140

 

 

 

OVP Trip Voltage

Model VMAX + 10%

OVSET 8.25

 

 

 

 

Delay

0.5 s

 

DLY .5S

 

 

 

 

Remote Enable

ON

 

REN ON

 

 

 

 

Foldback Protection

OFF

 

FOLD OFF

 

 

 

 

Output

ON

 

OUT ON

 

 

 

 

Hold

OFF

 

HOLD OFF

 

 

 

 

Unmask

NONE

 

UNMASK NONE

 

 

 

 

AUXA

OFF

 

AUXA OFF

 

 

 

 

AUXB

OFF

 

AUXB OFF

 

 

 

 

Remote Mode Operation

While in remote mode, use the interface commands to control the output of the power supply from computer. See “Operation” for a complete list of device-dependent commands available with this interface.

From remote mode, change to local mode operation by pressing the front panel LOCAL button or by sending either the GTL command or the REN OFF command. You can disable the LOCAL button by using the LLO command. For an example of how to use Local Lockout, see “Local Lockout (LLO) Command” on page 2–23.

To return to remote mode, ensure that the REN command is set to ON and then send any valid device-dependent command. Since the remote mode output settings may be different from the local mode settings, the output is programmed to protect the load by turning off when you toggle from local to remote mode during a session. Check your remote settings and then send OUT ON to restore the output.

TM-XR9B-01XN

2–21

Image 43
Contents Operating Manual Page XFR 1.2kW and 2.8kW Ethernet/RS-232 Interface Option Date and Revision Scope PurposeAudience Organization Related Information Conventions UsedPage Important Safety Instructions Viii Contents Operation Advanced Ethernet Administration SpecificationsCalibration Xii Figures Xiv Tables Xvi Description FeaturesInterface Features Features and FunctionsProgrammable Functions Ethernet FeaturesReadback Functions Page Introduction ConfigurationFor RS-232 Initial Inspection FLT Front PanelSRQ ERR1Item Ethernet/RS-232 Interface PCB Xoff 2Ethernet/RS-232 Interface Jumper SelectionsChanging Internal Jumpers Procedure for 1.2kW XFRProcedure for 2.8kW XFR Basic Setup Procedure Ethernet 1Step#DescriptionActionModeConfigure for Ethernet1Step Basic Setup Procedure RS-232Ethernet Configuring for Ethernet or RS-232Ethernet Connection Ethernet CommunicationsLocating the Power Supply on Your Network Configuring Ethernet Bridge Background Installing and Using RealPort1RS-232 Connection RS-232CommunicationsBaud Rate Selection 6Jumper J2 Settings for Baud Rate7Jumper J2 Flow Control Section Flow Control SelectionRemote/Local Operation Remote/Local Mode StartupRemote Mode Operation Remote Enable REN Command Local Mode OperationRTS/CTS Power Supply SettingsLocal Lockout LLO Command XON/XOFFOVP Selection Additional User Options and SettingsLowon TTL Shutdown PolarityHigh LOW OFF User Lines Connector User SignalsUser Lines Signal Connector Circuit Block Diagram 7User Lines Cable with Ferrite Block User Lines Cable ConnectionOperation RS-232Operation Command SyntaxManual Conventions Format Command Format and ParametersFlo ting Point Number floa Iset 2.0A Vset Command StringsCommand Terminators OrderCommand Summary ERR? AUXB?CMODE? DLY?ILO FAULT?Idata IHIAUXA? Command ReferenceCmode 1/ON Description AlsoRemote/Local Operation on Chapter Mask None = Unmask ALL ALL, None Command Voltage settings which are being held. See Hold command ASTS? Accumulated Status, Status, and Fault RegistersOPF PONREM ACFError Codes Error0 #ErrorNo ErrorsIdentificationExplanationDiagnostic LEDs Troubleshooting10 Ethernet Mode Tips Ethernet Mode Troubleshooting TipsPossibleCauses RS-232and/or Ethernet Mode Troubleshooting Tips RS-232Mode Troubleshooting TipsPage Calibration Calibration Voltage Calibration Setup Voltage Mode CalibrationVoltage Program Calibration Procedure Voltage Readback Calibration Procedure Current Calibration Setup Current Mode CalibrationCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Specifications OVP 40-30600-2 40-70 600-4 Page Administration Changing the Root Password Resetting the Configuration from a Browser Resetting the Configuration to DefaultsIntroduction Viewing Port Statistics and Settings Copying the Configuration to and from a ServerViewing Network Statistics NET Interpreting LED InformationDIA Page Warranty Warranty and Product InformationProduct DisclaimerExclusions Return Material Authorization PolicyOut of Warranty Service Return ProcedureInformation About Your System Page Xantrex Technology Inc