Calibration

You can calibrate the Ethernet/RS-232 Interface for:

Voltage program

Voltage readback

Current program

Current readback

Overvoltage protection

The following equipment will be required to accurately calibrate your unit:

Digital Voltmeter, 5 1/2 digit, 0.1% accuracy or better, with test leads

Current sensing shunt resistor, rated for 150% of maximum output current, 0.25% accuracy or better

Connection wires rated for the unit's maximum output current and voltage

A network ready, or RS-232 equipped computer to send the calibration commands to the unit

Calibrate the unit according to the following procedures, referring to “Command Reference” for more information about the calibration commands used.

The calibration procedures in this section are designed to be performed at an ambient temperature of 25°C ± 5°C.

4–2

TM-XR9B-01XN

Image 76
Contents Operating Manual Page XFR 1.2kW and 2.8kW Ethernet/RS-232 Interface Option Date and Revision Scope PurposeAudience Organization Conventions Used Related InformationPage Important Safety Instructions Viii Contents Operation Advanced Ethernet Administration SpecificationsCalibration Xii Figures Xiv Tables Xvi Features DescriptionFeatures and Functions Interface FeaturesProgrammable Functions Ethernet FeaturesReadback Functions Page Configuration IntroductionFor RS-232 Initial Inspection Front Panel SRQERR FLT1Item Ethernet/RS-232 Interface PCB 2Ethernet/RS-232 Interface Jumper Selections XoffProcedure for 1.2kW XFR Changing Internal JumpersProcedure for 2.8kW XFR 1Step#DescriptionActionModeConfigure for Ethernet Basic Setup Procedure EthernetBasic Setup Procedure RS-232 1StepConfiguring for Ethernet or RS-232 EthernetEthernet Communications Ethernet ConnectionLocating the Power Supply on Your Network Configuring Ethernet Bridge Installing and Using RealPort1 BackgroundRS-232Communications RS-232 Connection6Jumper J2 Settings for Baud Rate Baud Rate SelectionFlow Control Selection 7Jumper J2 Flow Control SectionRemote/Local Mode Startup Remote/Local OperationRemote Mode Operation Local Mode Operation Remote Enable REN CommandPower Supply Settings Local Lockout LLO CommandXON/XOFF RTS/CTSAdditional User Options and Settings OVP SelectionLowon TTL Shutdown PolarityHigh LOW OFF User Signals User Lines ConnectorUser Lines Signal Connector Circuit Block Diagram User Lines Cable Connection 7User Lines Cable with Ferrite BlockOperation RS-232Operation Command SyntaxManual Conventions Command Format and Parameters FormatFlo ting Point Number floa Command Strings Command TerminatorsOrder Iset 2.0A VsetCommand Summary AUXB? CMODE?DLY? ERR?FAULT? IdataIHI ILOAUXA? Command ReferenceCmode 1/ON Description AlsoRemote/Local Operation on Chapter Mask None = Unmask ALL ALL, None Command Voltage settings which are being held. See Hold command Accumulated Status, Status, and Fault Registers ASTS?PON REMACF OPFError0 #ErrorNo ErrorsIdentificationExplanation Error CodesTroubleshooting Diagnostic LEDsEthernet Mode Troubleshooting Tips 10 Ethernet Mode TipsPossibleCauses RS-232Mode Troubleshooting Tips RS-232and/or Ethernet Mode Troubleshooting TipsPage Calibration Calibration Voltage Calibration Setup Voltage Mode CalibrationVoltage Program Calibration Procedure Voltage Readback Calibration Procedure Current Calibration Setup Current Mode CalibrationCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Specifications 40-30 OVP600-2 40-70 600-4 Page Administration Changing the Root Password Resetting the Configuration from a Browser Resetting the Configuration to DefaultsIntroduction Viewing Port Statistics and Settings Copying the Configuration to and from a ServerViewing Network Statistics NET Interpreting LED InformationDIA Page Warranty and Product Information WarrantyDisclaimer ProductReturn Material Authorization Policy ExclusionsOut of Warranty Service Return ProcedureInformation About Your System Page Xantrex Technology Inc