Xantrex Technology ENET-XFR3 manual Voltage Mode Calibration, Voltage Calibration Setup

Page 77

Voltage Mode Calibration

Voltage Mode Calibration

Voltage Calibration Setup

1.Disconnect the load from the power supply which is to be calibrated.

2.Connect a voltmeter across the power supply’s output terminals.

Power Supply

Power Supply

Positive Output

Negative Output

 

 

 

 

 

 

 

 

 

 

 

 

V

+ -

VOLTMETER

Figure 4-1Voltage Calibration Setup

Voltage Program Calibration Procedure

1.Set the power supply for calibration as in Figure 4-1.

2.Activate calibration mode by sending command CMODE ON or CMODE 1 to the power supply.

3.Send command VLO; ILO to the power supply. Measure and record the output shown on the external voltmeter.

4.Send command VHI; IHI to the supply. Measure and record the output voltage as shown on the external voltmeter.

5.Send the command VDATA <vlo>,<vhi> where <vlo> and <vhi> are the values read from the voltmeter when the VLO and VHI commands were sent. When the power supply is calibrated, the low to high voltage program calibration values are stored as constants.

6.Program the supply at various levels using the VSET command to confirm that the calibration was successful and that linearity is observed. See the voltage program accuracy specification in Appendix A.

7.Turn off calibration mode by sending the command CMODE OFF or CMODE 0 to the power supply.

TM-XR9B-01XN

4–3

Image 77
Contents Operating Manual Page XFR 1.2kW and 2.8kW Ethernet/RS-232 Interface Option Date and Revision Audience PurposeScope Organization Related Information Conventions UsedPage Important Safety Instructions Viii Contents Operation Calibration SpecificationsAdvanced Ethernet Administration Xii Figures Xiv Tables Xvi Description FeaturesInterface Features Features and FunctionsReadback Functions Ethernet FeaturesProgrammable Functions Page Introduction ConfigurationFor RS-232 Initial Inspection SRQ Front PanelERR FLT1Item Ethernet/RS-232 Interface PCB Xoff 2Ethernet/RS-232 Interface Jumper SelectionsChanging Internal Jumpers Procedure for 1.2kW XFRProcedure for 2.8kW XFR Basic Setup Procedure Ethernet 1Step#DescriptionActionModeConfigure for Ethernet1Step Basic Setup Procedure RS-232Ethernet Configuring for Ethernet or RS-232Ethernet Connection Ethernet CommunicationsLocating the Power Supply on Your Network Configuring Ethernet Bridge Background Installing and Using RealPort1RS-232 Connection RS-232CommunicationsBaud Rate Selection 6Jumper J2 Settings for Baud Rate7Jumper J2 Flow Control Section Flow Control SelectionRemote/Local Operation Remote/Local Mode StartupRemote Mode Operation Remote Enable REN Command Local Mode OperationLocal Lockout LLO Command Power Supply SettingsXON/XOFF RTS/CTSOVP Selection Additional User Options and SettingsHigh LOW OFF TTL Shutdown PolarityLowon User Lines Connector User SignalsUser Lines Signal Connector Circuit Block Diagram 7User Lines Cable with Ferrite Block User Lines Cable ConnectionOperation Manual Conventions Command SyntaxRS-232Operation Format Command Format and ParametersFlo ting Point Number floa Command Terminators Command StringsOrder Iset 2.0A VsetCommand Summary CMODE? AUXB?DLY? ERR?Idata FAULT?IHI ILOCmode 1/ON Command ReferenceAUXA? Description AlsoRemote/Local Operation on Chapter Mask None = Unmask ALL ALL, None Command Voltage settings which are being held. See Hold command ASTS? Accumulated Status, Status, and Fault RegistersREM PONACF OPFError Codes Error0 #ErrorNo ErrorsIdentificationExplanationDiagnostic LEDs Troubleshooting10 Ethernet Mode Tips Ethernet Mode Troubleshooting TipsPossibleCauses RS-232and/or Ethernet Mode Troubleshooting Tips RS-232Mode Troubleshooting TipsPage Calibration Calibration Voltage Program Calibration Procedure Voltage Mode CalibrationVoltage Calibration Setup Voltage Readback Calibration Procedure Current Program Calibration Procedure Current Mode CalibrationCurrent Calibration Setup Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Specifications OVP 40-30600-2 40-70 600-4 Page Administration Changing the Root Password Introduction Resetting the Configuration to DefaultsResetting the Configuration from a Browser Viewing Network Statistics Copying the Configuration to and from a ServerViewing Port Statistics and Settings DIA Interpreting LED InformationNET Page Warranty Warranty and Product InformationProduct DisclaimerExclusions Return Material Authorization PolicyInformation About Your System Return ProcedureOut of Warranty Service Page Xantrex Technology Inc