Page 59

Command Reference

Command Reference

Table 3-7Command Reference

ASTS?Command

DescriptionAsks for the supply’s accumulated status register. The accumulated status register

 

stores any bit that was entered in the status register since the accumulated status

 

query command (ASTS?) was last used, regardless of whether the condition still

 

exists. The accumulated status register has the same bits, weights, and conditions

 

as the status register. A bit in the accumulated status register will be set at 1 if the

 

corresponding bit in the status register has been 1 (TRUE) at any time since the

 

register was last read. See “Accumulated Status, Status, and Fault Registers” on

 

page 3–17. The ASTS? query clears the status register.

 

Response: ASTS <status mask> where status mask is the decimal equivalent of the

 

total bit weights for the operating conditions as listed in the status register.

 

 

AUXA <1/ON>,

Controls the AUXA output signal level at rear panel connector Pin 7. Active low.

<0/OFF>

Initial value: AUXA 0

 

 

AUXA?

Asks for the present set value of the AUXA output signal.

 

Response: AUXA 0 (OFF) or AUXA 1 (ON)

 

 

AUXB <1/ON>,

Controls the AUXB output signal level at rear panel connector Pin 8. Active low.

<0/OFF>

Initial value: AUXB 0

 

 

AUXB?

Asks for the present set value of the AUXB output signal.

 

Response: AUXB 0 (OFF) or AUXB 1 (ON)

 

 

CLR

Initializes the power supply to its power ON condition. If issued while in local

 

mode, CLR will force power supply settings to register default values as in but

 

these default settings will not come into effect until the power supply is switched to

 

remote mode operation. The CLR commands will clear faults from the fault

 

register. CLR will not reset CMODE.

 

 

CMODE <1/ON>,

CMODE ON places the power supply into calibration mode for processing

<0/OFF>

calibration commands.

 

Initial value: CMODE OFF or CMODE 0

 

 

CMODE?

Asks for the power supply’s calibration mode status.

 

Response: CMODE 0 (disabled)

 

CMODE 1 (enabled)

 

 

TM-XR9B-01XN

3–9

Image 59
Contents Operating Manual Page XFR 1.2kW and 2.8kW Ethernet/RS-232 Interface Option Date and Revision Audience PurposeScope Organization Related Information Conventions UsedPage Important Safety Instructions Viii Contents Operation Calibration SpecificationsAdvanced Ethernet Administration Xii Figures Xiv Tables Xvi Description FeaturesInterface Features Features and FunctionsReadback Functions Ethernet FeaturesProgrammable Functions Page Introduction ConfigurationFor RS-232 Initial Inspection FLT Front PanelSRQ ERR1Item Ethernet/RS-232 Interface PCB Xoff 2Ethernet/RS-232 Interface Jumper SelectionsChanging Internal Jumpers Procedure for 1.2kW XFRProcedure for 2.8kW XFR Basic Setup Procedure Ethernet 1Step#DescriptionActionModeConfigure for Ethernet1Step Basic Setup Procedure RS-232Ethernet Configuring for Ethernet or RS-232Ethernet Connection Ethernet CommunicationsLocating the Power Supply on Your Network Configuring Ethernet Bridge Background Installing and Using RealPort1RS-232 Connection RS-232CommunicationsBaud Rate Selection 6Jumper J2 Settings for Baud Rate7Jumper J2 Flow Control Section Flow Control SelectionRemote/Local Operation Remote/Local Mode StartupRemote Mode Operation Remote Enable REN Command Local Mode OperationRTS/CTS Power Supply SettingsLocal Lockout LLO Command XON/XOFFOVP Selection Additional User Options and SettingsHigh LOW OFF TTL Shutdown PolarityLowon User Lines Connector User SignalsUser Lines Signal Connector Circuit Block Diagram 7User Lines Cable with Ferrite Block User Lines Cable ConnectionOperation Manual Conventions Command SyntaxRS-232Operation Format Command Format and ParametersFlo ting Point Number floa Iset 2.0A Vset Command StringsCommand Terminators OrderCommand Summary ERR? AUXB?CMODE? DLY?ILO FAULT?Idata IHICmode 1/ON Command ReferenceAUXA? Description AlsoRemote/Local Operation on Chapter Mask None = Unmask ALL ALL, None Command Voltage settings which are being held. See Hold command ASTS? Accumulated Status, Status, and Fault RegistersOPF PONREM ACFError Codes Error0 #ErrorNo ErrorsIdentificationExplanationDiagnostic LEDs Troubleshooting10 Ethernet Mode Tips Ethernet Mode Troubleshooting TipsPossibleCauses RS-232and/or Ethernet Mode Troubleshooting Tips RS-232Mode Troubleshooting TipsPage Calibration Calibration Voltage Program Calibration Procedure Voltage Mode CalibrationVoltage Calibration Setup Voltage Readback Calibration Procedure Current Program Calibration Procedure Current Mode CalibrationCurrent Calibration Setup Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Specifications OVP 40-30600-2 40-70 600-4 Page Administration Changing the Root Password Introduction Resetting the Configuration to DefaultsResetting the Configuration from a Browser Viewing Network Statistics Copying the Configuration to and from a ServerViewing Port Statistics and Settings DIA Interpreting LED InformationNET Page Warranty Warranty and Product InformationProduct DisclaimerExclusions Return Material Authorization PolicyInformation About Your System Return ProcedureOut of Warranty Service Page Xantrex Technology Inc