Xantrex Technology ENET-XFR3 manual Configuring Ethernet Bridge

Page 37

Ethernet Communications

Configuring Ethernet Bridge

The Ethernet/RS-232 Interface card has an Ethernet Bridge that converts the ethernet packets to serial data signals for processing by the embedded microcontroller. The Ethernet Bridge must be properly configured for serial communications with the microcontroller. For example, settings on the Bridge, such as Baud Rate and Flow Control, must match the jumper settings on the interface board.

There are two ways to gain access to the configuration menu for the Ethernet Bridge. One is to run the Setup program that is available on the Software CD. This program will find your instrument and allow you to configure it by clicking on the instrument’s IP address. You will then need to enter the correct username and password to be able to configure the device.

Default Username: root

Default Password: dbps

The other method is to open up a web browser like Internet Explorer, and type in the IP address for the instrument. Again, you will be asked for the username and password. Once you have accessed the configuration page, follow these steps for proper setup.

1.Click Configuration > Serial Port, and ensure that the following settings have been selected:

Baud Rate:

9600

Data Bits:

8

Parity:

None

Stop Bits:

1

Flow Control:

None

Click Apply once you have finished adjusting the above settings.

2.Now click Configuration > Network. If your network uses a DHCP server (ask your IT department) you should select Obtain IP address automatically using DHCP and then click Apply. If your network doesn’t use a DHCP server, select Use the following IP address: and enter an appropriate IP address, Subnet Mask and Default Gateway. .

3.Ask for assistance from your IT department to obtain an appropriate IP address, Subnet Mask, Default Gateway, Name Server, and Domain. Once you have made the appropriate changes, click Apply. If this changed the setting, the device will reboot, so you will have to renew your connection to its configuration webpage for any further configuration changes.

TM-XR9B-01XN

2–15

Image 37
Contents Operating Manual Page XFR 1.2kW and 2.8kW Ethernet/RS-232 Interface Option Date and Revision Scope PurposeAudience Organization Related Information Conventions UsedPage Important Safety Instructions Viii Contents Operation Advanced Ethernet Administration SpecificationsCalibration Xii Figures Xiv Tables Xvi Description FeaturesInterface Features Features and FunctionsProgrammable Functions Ethernet FeaturesReadback Functions Page Introduction ConfigurationFor RS-232 Initial Inspection SRQ Front PanelERR FLT1Item Ethernet/RS-232 Interface PCB Xoff 2Ethernet/RS-232 Interface Jumper SelectionsChanging Internal Jumpers Procedure for 1.2kW XFRProcedure for 2.8kW XFR Basic Setup Procedure Ethernet 1Step#DescriptionActionModeConfigure for Ethernet1Step Basic Setup Procedure RS-232Ethernet Configuring for Ethernet or RS-232Ethernet Connection Ethernet CommunicationsLocating the Power Supply on Your Network Configuring Ethernet Bridge Background Installing and Using RealPort1RS-232 Connection RS-232CommunicationsBaud Rate Selection 6Jumper J2 Settings for Baud Rate7Jumper J2 Flow Control Section Flow Control SelectionRemote/Local Operation Remote/Local Mode StartupRemote Mode Operation Remote Enable REN Command Local Mode OperationLocal Lockout LLO Command Power Supply SettingsXON/XOFF RTS/CTSOVP Selection Additional User Options and SettingsLowon TTL Shutdown PolarityHigh LOW OFF User Lines Connector User SignalsUser Lines Signal Connector Circuit Block Diagram 7User Lines Cable with Ferrite Block User Lines Cable ConnectionOperation RS-232Operation Command SyntaxManual Conventions Format Command Format and ParametersFlo ting Point Number floa Command Terminators Command StringsOrder Iset 2.0A VsetCommand Summary CMODE? AUXB?DLY? ERR?Idata FAULT?IHI ILOAUXA? Command ReferenceCmode 1/ON Description AlsoRemote/Local Operation on Chapter Mask None = Unmask ALL ALL, None Command Voltage settings which are being held. See Hold command ASTS? Accumulated Status, Status, and Fault RegistersREM PONACF OPFError Codes Error0 #ErrorNo ErrorsIdentificationExplanationDiagnostic LEDs Troubleshooting10 Ethernet Mode Tips Ethernet Mode Troubleshooting TipsPossibleCauses RS-232and/or Ethernet Mode Troubleshooting Tips RS-232Mode Troubleshooting TipsPage Calibration Calibration Voltage Calibration Setup Voltage Mode CalibrationVoltage Program Calibration Procedure Voltage Readback Calibration Procedure Current Calibration Setup Current Mode CalibrationCurrent Program Calibration Procedure Current Readback Calibration Procedure Over Voltage Protection OVP Calibration Page Specifications OVP 40-30600-2 40-70 600-4 Page Administration Changing the Root Password Resetting the Configuration from a Browser Resetting the Configuration to DefaultsIntroduction Viewing Port Statistics and Settings Copying the Configuration to and from a ServerViewing Network Statistics NET Interpreting LED InformationDIA Page Warranty Warranty and Product InformationProduct DisclaimerExclusions Return Material Authorization PolicyOut of Warranty Service Return ProcedureInformation About Your System Page Xantrex Technology Inc