Sierra 241-VTP, 240-VTP instruction manual Unobstructed Flow Requirements

Page 16

Chapter 2 Installation

Series 24-HP Instruction Manual

 

 

Unobstructed Flow Requirements

Select an installation site that will minimize possible distortion in the flow profile. Valves, elbows, control valves and other piping components may cause flow disturbances. Check your specific piping condition against the examples shown below. In order to achieve accurate and re- peatable performance install the flow meter using the recommended number of straight run pipe diameters upstream and downstream of the sensor.

Note: For liquid applications in vertical pipes, avoid installing with flow in the downward direction because the pipe may not be full at all points. Choose to install the meter with flow in the upward direction if possible.

A

B

A

B

Flow meter

Example 1.

One 90°elbow before meter

C'

C

Flow meter

 

 

Example 4.

Flow conditioner

 

(if used)

Reduction before meter

 

 

 

AB

C'

 

 

C

Flow meter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow conditioner (if used)

Example 2.

Two 90°elbows before meter in one plane

AB

C'

C

Flow meter

 

 

Example 5.

Flow conditioner

Expansion before meter

(if used)

AB

C'

 

 

C

Flow meter

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Flow conditioner (if used)

Example 3.

Two 90°elbows before meter out of plane (if three 90°bends present, double recommended length)

A

 

B

C'

C

Flow meter

 

 

Flow conditioner (if used)

Example 6.

Regulator or valve partially closed before meter (If valve is always wide open, base length requirements on fitting directly preceding it)

 

 

Minimum Required

 

 

Minimum Required

 

 

Upstream Diameters

 

 

Downstream Diameters

 

 

No Flow

 

 

 

 

No Flow

With Flow

 

 

Conditioner

With Flow Conditioner

Conditioner

Conditioner

 

Example

A

A

 

C

B

B

 

1

10 D

N/A

 

N/A

N/A

5 D

5 D

 

2

15 D

10 D

 

5 D

5 D

5 D

5 D

 

3

25 D

10 D

 

5 D

5 D

10 D

5 D

 

4

10 D

10 D

 

5 D

5 D

5 D

5 D

 

5

20 D

10 D

 

5 D

5 D

5 D

5 D

 

6

25 D

10 D

 

5 D

5 D

10 D

5 D

 

 

D = Internal diameter of channel. N/A = Not applicable

 

 

 

 

 

Figure 2-1. Recommended Pipe Length Requirements for Installation, Series 240 and 241

 

 

 

 

 

 

 

 

 

 

2-2

 

 

 

 

 

 

 

IM-24-HP

Image 16
Contents Models 240-VT, 240-VTP, 241-VT, 241-VTP Customer Notice Copyright Sierra InstrumentsTable of Contents Troubleshooting and Repair Figures TablesTable of Contents Introduction Using This ManualInnova-MassMulti-Parameter Vortex Mass Flow Meters Receipt of System Components Technical AssistanceHow the Innova-Mass Vortex Mass Flow Meter Operates Velocity MeasurementVortex Shedding Frequency Vortex Frequency SensingFlow Velocity Range Gas LiquidTemperature Measurement Reynolds Number Range for the Innova-MassFlow Meter Configurations Pressure MeasurementFlow Meter Electronics Installation Overview InstallationFlow Meter Installation Requirements Unobstructed Flow Requirements Series 240 In-Line Flow Meter Installation Flange Bolt SpecificationsWafer-Style Flow Meter Installation Wafer-Style Flow Meter InstallationFlange-Style Flow Meter Installation Flange-Style Flow Meter InstallationMounting Position Series 241 Insertion Flow Meter InstallationIsolation Valve Selection Cold Tap Guidelines Hot Tap Guidelines FlowFlow Meter Insertion Use the Correct Insertion FormulaInstalling Flow Meters with a Compression Connection ExampleInsertion Procedure for Meters with a Compression Connection Example 1 Flange Style Meters Installing Flow Meters with a Packing Gland ConnectionExample 2 NPT Style Meters Flow Meter with Permanent Insertion Tool 10. Flow Meter with Removable Insertion Tool 11. Insertion Calculation Meters without Insertion Tool Installation Adjusting Meter Orientation Display/Keypad Adjustment All MetersEnclosure Adjustment Series 240 Only 13. Enclosure Viewing AdjustmentWiring Connections Input Power ConnectionsAC Power Wiring DC Power WiringMA Output Connections 16. Load Resistance Versus Input VoltagePulse Output Connections 20. Isolated Pulse Output with External Power Supply Alarm Output Connections 23. Isolated Alarm Output with External Power SupplyRemote Electronics Wiring 25. Isolated Alarm Output Using Meter Provided Power SupplyOptional Input Electronics Wiring Optional Energy EMS Input Electronics WiringPage Flow Meter Display/Keypad Operating InstructionsSierra Run Mode Screens Start-UpUsing the Setup Menus Programming the Flow MeterOutput Menu TimeConstSec XxxxExample for Setting an Output First, set the desired units of measurementExample for Changing a Run Mode Display Item Display MenuAlarms Menu Example for Setting an AlarmExample for Setting the Totalizer Totalizer #1 MenuTotalizer #2 Menu Configuration Energy Menu For EMS Energy Meters OnlyFluid Menu Units Menu Example for Setting the Time Time & Date MenuDiagnostics Menu Calibration Menu Password Menu Troubleshooting and Repair Hidden Diagnostics MenusColumn One Hidden Diagnostics Values IM-24-HP A2D Ref. Resistor = Factory Use Only Analog Output Calibration Troubleshooting the Flow Meter Symptom Output at no Flow Symptom No Output Symptom Meter Displays Temperature Fault Electronics Assembly Replacement All Meters Pressure Sensor Replacement Appendix a Product Specifications Appendix a Specifications Appendix a Specifications Cnpt IP66, Nema Page Appendix B Glossary C DK L S T Page Appendix C Fluid Calculations Calculations for Steam T & PDensity ViscosityCalculations for Gas Real Gas and Other Gas Density for real gases is calculated from the equationCalculations for Liquid Page Appendix D Hart Commands with DD Menu PV AOAppendix D Hart Commands without DD Menu PV AO PV LRV URVAppendix E Modbus Communication Applicable Flow Meter ModelsModbus Order Register Definitions Appendix E Modbus Commands Following registers contain the display units strings Control Register Definitions Examples An attempt to read registers that don’t exist 8C 3A CRC not the correct CRC EJS-02-06-07 Appendix E Modbus Commands