Sierra 240-VT Appendix C Fluid Calculations, Calculations for Steam T & P, Density, Viscosity

Page 77

Series 24-HP Instruction Manual

Appendix C Fluid Calculations

 

 

Appendix C Fluid Calculations

Calculations for Steam T & P

When “Steam T & P” is selected in the “Real Gas” selection of the Fluid

Menu, the calculations are based on the equations below.

Density

The density of steam is calculated from the formula given by Keenan and Keys. The given equation is for the volume of the steam.

v = 4.555.04 T + B

p

B= B0 + B0 2 g1 (τ )τ ⋅ p + B0 4 g 2 (τ )τ 3 p 3 B013 g 3 (τ )τ 12 p12

B0 = 1.89 2641.62 ⋅τ ⋅1080870τ 2

g

(τ ) = 82.546 ⋅τ −1.6246 105 ⋅τ 2

1

 

g2 (τ ) = 0.21828 1.2697 105 ⋅τ 2

g3 (τ ) = 3.635 104 6.7681064 ⋅τ 24

Where tau is 1/ temperature in Kelvin.

The density can be found from 1/(v/ standard density of water).

Viscosity

The viscosity is based on an equation given by Keenan and Keys.

η( poise) = 1.501105 T 1 + 446.8 / T

Where T is the temperature in Kelvin

IM-24-HP

C-1

Image 77
Contents Models 240-VT, 240-VTP, 241-VT, 241-VTP Copyright Sierra Instruments Customer NoticeTable of Contents Troubleshooting and Repair Tables FiguresTable of Contents Innova-MassMulti-Parameter Vortex Mass Flow Meters Using This ManualIntroduction Technical Assistance Receipt of System ComponentsVelocity Measurement How the Innova-Mass Vortex Mass Flow Meter OperatesVortex Frequency Sensing Vortex Shedding FrequencyGas Liquid Flow Velocity RangeReynolds Number Range for the Innova-Mass Temperature MeasurementPressure Measurement Flow Meter ConfigurationsFlow Meter Electronics Flow Meter Installation Requirements InstallationInstallation Overview Unobstructed Flow Requirements Flange Bolt Specifications Series 240 In-Line Flow Meter InstallationWafer-Style Flow Meter Installation Wafer-Style Flow Meter InstallationFlange-Style Flow Meter Installation Flange-Style Flow Meter InstallationIsolation Valve Selection Series 241 Insertion Flow Meter InstallationMounting Position Cold Tap Guidelines Flow Hot Tap GuidelinesUse the Correct Insertion Formula Flow Meter InsertionExample Installing Flow Meters with a Compression ConnectionInsertion Procedure for Meters with a Compression Connection Example 2 NPT Style Meters Installing Flow Meters with a Packing Gland ConnectionExample 1 Flange Style Meters Flow Meter with Permanent Insertion Tool 10. Flow Meter with Removable Insertion Tool 11. Insertion Calculation Meters without Insertion Tool Installation Display/Keypad Adjustment All Meters Adjusting Meter Orientation13. Enclosure Viewing Adjustment Enclosure Adjustment Series 240 OnlyInput Power Connections Wiring ConnectionsAC Power Wiring DC Power Wiring16. Load Resistance Versus Input Voltage MA Output ConnectionsPulse Output Connections 20. Isolated Pulse Output with External Power Supply 23. Isolated Alarm Output with External Power Supply Alarm Output Connections25. Isolated Alarm Output Using Meter Provided Power Supply Remote Electronics WiringOptional Energy EMS Input Electronics Wiring Optional Input Electronics WiringPage Sierra Operating InstructionsFlow Meter Display/Keypad Start-Up Run Mode ScreensProgramming the Flow Meter Using the Setup MenusTimeConstSec Xxxx Output MenuFirst, set the desired units of measurement Example for Setting an OutputDisplay Menu Example for Changing a Run Mode Display ItemExample for Setting an Alarm Alarms MenuTotalizer #1 Menu Example for Setting the TotalizerTotalizer #2 Menu Energy Menu For EMS Energy Meters Only ConfigurationFluid Menu Units Menu Time & Date Menu Example for Setting the TimeDiagnostics Menu Calibration Menu Password Menu Hidden Diagnostics Menus Troubleshooting and RepairColumn One Hidden Diagnostics Values IM-24-HP A2D Ref. Resistor = Factory Use Only Analog Output Calibration Troubleshooting the Flow Meter Symptom Output at no Flow Symptom No Output Symptom Meter Displays Temperature Fault Electronics Assembly Replacement All Meters Pressure Sensor Replacement Appendix a Product Specifications Appendix a Specifications Appendix a Specifications Cnpt IP66, Nema Page C D Appendix B GlossaryK L S T Page Calculations for Steam T & P Appendix C Fluid CalculationsDensity ViscosityDensity for real gases is calculated from the equation Calculations for Gas Real Gas and Other GasCalculations for Liquid Page PV AO Appendix D Hart Commands with DD MenuPV AO PV LRV URV Appendix D Hart Commands without DD MenuApplicable Flow Meter Models Appendix E Modbus CommunicationModbus Order Register Definitions Appendix E Modbus Commands Following registers contain the display units strings Control Register Definitions Examples An attempt to read registers that don’t exist 8C 3A CRC not the correct CRC EJS-02-06-07 Appendix E Modbus Commands