Chapter 3 Operation

 

 

Example for Setting an Output

The following shows how to set Output 1 to measure mass flow with 4 mA = 0 lb/hr and 20 mA = 100 lb/hr with a time constant of 5 seconds. (All outputs are disabled while using the Setup Menus.)

First, set the desired units of measurement:

1.Use ￿￿ keys to move to the Units Menu (see page 3-9).

2.Press ￿ key until Mass Flow Unit appears. Press ENTER.

3.Press ￿ key until lb appears in the numerator. Press ￿ key to move the underline cursor to the denominator. Press the ￿ key until hr appears in the denominator. Press ENTER to select.

4.Press ￿ key until Units Menu appears.

Second, set the analog output:

1.Use ￿￿ keys to move to the Output Menu.

2.Press the ￿ key until 4-20 mA Output 1 appears.

3.Press ￿ key to access Measure selections. Press ENTER and press the ￿ key to select Mass. Press ENTER.

4.Press ￿ key to set the 4 mA point in the units you have selected for mass of lb/hr. Press ENTER and use ￿￿￿￿ keys to set 0 or 0.0. Press ENTER.

5.Press ￿ key to set the 20 mA point. Press ENTER and use ￿￿￿￿ keys to set 100 or 100.0. Press ENTER.

6.Press ￿ key to select the Time Constant. Press ENTER and use ￿￿￿￿ keys to select 5. Press ENTER.

7.Press the EXIT key and answer YES to permanently save your changes.

IM-24-HP

3-5

Image 45
Contents Models 240-VT, 240-VTP, 241-VT, 241-VTP Copyright Sierra Instruments Customer NoticeTable of Contents Troubleshooting and Repair Tables FiguresTable of Contents Using This Manual IntroductionInnova-MassMulti-Parameter Vortex Mass Flow Meters Technical Assistance Receipt of System ComponentsVelocity Measurement How the Innova-Mass Vortex Mass Flow Meter OperatesVortex Frequency Sensing Vortex Shedding FrequencyGas Liquid Flow Velocity RangeReynolds Number Range for the Innova-Mass Temperature MeasurementPressure Measurement Flow Meter ConfigurationsFlow Meter Electronics Installation Installation OverviewFlow Meter Installation Requirements Unobstructed Flow Requirements Flange Bolt Specifications Series 240 In-Line Flow Meter InstallationWafer-Style Flow Meter Installation Wafer-Style Flow Meter InstallationFlange-Style Flow Meter Installation Flange-Style Flow Meter InstallationSeries 241 Insertion Flow Meter Installation Mounting PositionIsolation Valve Selection Cold Tap Guidelines Flow Hot Tap GuidelinesUse the Correct Insertion Formula Flow Meter InsertionExample Installing Flow Meters with a Compression ConnectionInsertion Procedure for Meters with a Compression Connection Installing Flow Meters with a Packing Gland Connection Example 1 Flange Style MetersExample 2 NPT Style Meters Flow Meter with Permanent Insertion Tool 10. Flow Meter with Removable Insertion Tool 11. Insertion Calculation Meters without Insertion Tool Installation Display/Keypad Adjustment All Meters Adjusting Meter Orientation13. Enclosure Viewing Adjustment Enclosure Adjustment Series 240 OnlyInput Power Connections Wiring ConnectionsAC Power Wiring DC Power Wiring16. Load Resistance Versus Input Voltage MA Output ConnectionsPulse Output Connections 20. Isolated Pulse Output with External Power Supply 23. Isolated Alarm Output with External Power Supply Alarm Output Connections25. Isolated Alarm Output Using Meter Provided Power Supply Remote Electronics WiringOptional Energy EMS Input Electronics Wiring Optional Input Electronics WiringPage Operating Instructions Flow Meter Display/KeypadSierra Start-Up Run Mode ScreensProgramming the Flow Meter Using the Setup MenusTimeConstSec Xxxx Output MenuFirst, set the desired units of measurement Example for Setting an OutputDisplay Menu Example for Changing a Run Mode Display ItemExample for Setting an Alarm Alarms MenuTotalizer #1 Menu Example for Setting the TotalizerTotalizer #2 Menu Energy Menu For EMS Energy Meters Only ConfigurationFluid Menu Units Menu Time & Date Menu Example for Setting the TimeDiagnostics Menu Calibration Menu Password Menu Hidden Diagnostics Menus Troubleshooting and RepairColumn One Hidden Diagnostics Values IM-24-HP A2D Ref. Resistor = Factory Use Only Analog Output Calibration Troubleshooting the Flow Meter Symptom Output at no Flow Symptom No Output Symptom Meter Displays Temperature Fault Electronics Assembly Replacement All Meters Pressure Sensor Replacement Appendix a Product Specifications Appendix a Specifications Appendix a Specifications Cnpt IP66, Nema Page C D Appendix B GlossaryK L S T Page Calculations for Steam T & P Appendix C Fluid CalculationsDensity ViscosityDensity for real gases is calculated from the equation Calculations for Gas Real Gas and Other GasCalculations for Liquid Page PV AO Appendix D Hart Commands with DD MenuPV AO PV LRV URV Appendix D Hart Commands without DD MenuApplicable Flow Meter Models Appendix E Modbus CommunicationModbus Order Register Definitions Appendix E Modbus Commands Following registers contain the display units strings Control Register Definitions Examples An attempt to read registers that don’t exist 8C 3A CRC not the correct CRC EJS-02-06-07 Appendix E Modbus Commands