Chapter 3 Operation | Series |
|
|
Display Menu
ENTER
Run Mode | Password |
ENTER
Display
Menu
Cycle Time(Sec) 0
Number of Digits 2
Display TC(sec) 1
Use keys to access menus
If Cycle Time is set to zero, manual advance is required
Used to set t he number of digits displayed after decimal point
TC = Display Time Constant, used to smooth display
MF Vf Te Pr De T Y or N
A1 A2 A3 Fl Dt E
Y or N
Mf = Mass Flow
Vf = Volume Flow
Te = Temperature
Pr = Pressure
De = Density
T = Total
A1 = Alarm 1 Status
A2 = Alarm 2 Status
A3 = Alarm 3 Status
Fl = Fluid
Dt = Density
* E = Energy
For each parameter:
Select Yes to view parameter in Run Mode Select No to hide parameter in Run Mode
*Energy EMS Meters Only
Use the Display Menu to set the cycle time for automatic screen sequencing used in the Run Mode, change the precision of displayed values, smooth the values or enable or disable each item displayed in the Run Mode screens.
Example for Changing a Run Mode Display Item
The following shows how to remove the temperature screen from the Run Mode screens. Note: all outputs are disabled while using the Setup Menus.
1.Use keys to move to the Display Menu.
2.Press key until Mf Vf Pr Te De T appears.
3.Press ENTER to select. Press key until the cursor is positioned below Te.
4.Press key until N appears. Press ENTER to select.
5.Press EXIT and then ENTER to save changes and return to the Run Mode.