Figure 2-1. Power Module Line Fuse and Switches
| Figure |
| |||||||
Table |
| ||||||||
|
|
|
| Required Switch Setting |
|
| |||
Desired Function |
| 1 | 2 | 3 | 4 | 5 | 6 | 7* | 8* |
RI disabled |
| 0 | 0 |
|
|
|
| 1 | 1 |
RI unlatched (live) |
| 0 | 1 |
|
|
|
| 1 | 1 |
RI latching † |
| 1 | 1 |
|
|
|
| 1 | 1 |
Display disabled |
|
|
| 0 |
|
|
| 1 | 1 |
Display enabled † |
|
|
| I |
|
|
| 1 | 1 |
Inhibit calibration |
|
|
|
| 0 | 0 |
| 1 | 1 |
Factory calibration |
|
|
|
| 0 | 1 |
| 1 | 1 |
Normal calibration † |
|
|
|
| 1 | 1 |
| 1 | 1 |
Power on state is *RCL 0 |
|
|
|
|
| 0 | 1 | 1 | |
Power on state is *RST† |
|
|
|
|
|
| 1 | 1 | 1 |
†= factory default setting | "0" Off or open | "1" On or closed. |
|
|
|
|
*Positions 7 and 8 are for service functions. For normal operation, they must be set to 1.