Agilent Technologies 66001A manual Installing The Module In The Mainframe, Connecting The Load

Page 20

Installing The Module In The Mainframe

 

Figure 2-3. Power Module Installation

 

 

Note

A fully loaded mainframe can weigh over 36 kg (80 lbs). Install the mainframe into the rack (see the

 

"Mainframe Users Guide") before installing the modules into the mainframe.

 

 

1.Turn off the mainframe power switch.

2.Make certain the pull tab extends from the module (see Figure 2-3).

3.Slide the module into the mainframe slot. You may use any available slot.

4.Push the module into the backplane connector until you feel the module connector snap into the backplane connector.

5.Push the lock tab into the front of the module. If the module is not properly plugged into the backplane connector, the lock tab will not retract into the module.

6.If desired, mark or label the power module with its slot address.

Important Marking or labeling the power module will prevent replacing it in the wrong slot if it is subsequently removed from the mainframe. Also, with two mainframes linked together, a slot can be either of two numbers, depending on whether it is in the main or the auxiliary mainframe (see "Connecting the Controller" in this chapter).

Connecting The Load

Output Connector

One output connector (see Table 1-3) is shipped with each power module. This can be either a standard or relay type. The Installation Guide supplied with the connector shows how to open the connector, connect the wires, and secure the connector to the mainframe.

20 Installation

Image 20
Contents USER’S Guide Agilent Part No Microfiche Part No SeptemberCertification Safety Summary Symbol Description Printing HistoryHerstellerbescheinigung Manufacturers DeclarationModel Numbers a 66000A Manufacturer’s Name and AddressUSA Page Table of Contents Basic Power Module Commands Using the Agilent 66001A MPS KeyboardGeneral Information Item DescriptionOption Description Accessories Specifications and Supplemental Characteristics Programming Accuracy @ calibration temperature ± 5 COutput Ratings Readback AccuracyAgilent Model 66101A 66102A 66103A 66104A 66105A 66106A Performance Specifications ParameterSupplemental Characteristics Parameter 66101A 66102A Agilent Model 66103A 66104A 66105A 66106A Remote Sensing CapabilityRecommended Calibration Interval Safety Compliance DimensionsWeight Downprogramming Monotonicity Gpib Interface CapabilitiesSupplemental Characteristics Output Characteristic CurveReplaceable Parts List Description Agilent Part No Parameter Agilent Model Number and Parameter ValuePage Inspection InstallationSwitches Power Module Line Fuse and Switches Remote Inhibit RI Function Switches Switch Setting RI ActionDisplay Function Switch Calibration Function SwitchesOutput Connector Installing The Module In The MainframeConnecting The Load Selecting the Proper Wire Size Output ConnectionsResistance Ω/ft AWG No. *AmpacityRemote Voltage Sensing Voltage ReadbackParallel Operation Output RatingOutput Noise Outp OFF VoltlevVoltprot Currlev Voltlev Voltprot Currlev MAX Outp onMultiple Loads Capacitive LoadsConnecting The Controller Battery ChargingInductive Loads OVP ConsiderationsPower Module Front Panel Display Turn-On CheckoutIntroduction Module Panel DisplayAddr Dis Prot Power Module Turn-On StateChecking Basic Module Functions Module Output Connector Error MessagesCase Of Trouble Mainframe FailureSelftest Error Messages Runtime Error MessagesMessages DisplayBasic Power Module Commands Synopsis Of CommandsGpib Controller MPS KeyboardCommand Power Module Output Commands FunctionPower Module Protection Commands Function Power Module Fixed-Mode Output CommandsPower Module List-Mode Output Commands Power Module Trigger Commands FunctionControlling List-Mode Output Controlling TriggersChanging The Power On State Using the RI/DFI Functions RI Remote Inhibit InputDFI Discrete Fault Indicator Output Command DFI Output Commands FunctionChanging The Calibration Password CalibrationEnabling Or Disabling The Calibration Function Performing a CalibrationTable A-2. Voltage Calibration Procedure Equipment RequiredCalibrating Voltage Keyboard Entry Controller Entry † FunctionTable A-3. Current Calibration Procedure Disable the outputCalcurrlev MIN Calcurrlev MAXRecovering The Factory Calibration Constants Recovering From a Lost Calibration PasswordCalibration Error Messages Agilent Basic Calibration ProgramAssign Power Supply Gpib Address Initialize Power SupplyPassword is Optional only 10 ! Voltage Calibration ProgramEND if Print Current Calibration not SavedElse Print Current Calibration CompleteAgilent 66001A MPS Keyboard Using The Agilent 66001A MPS KeyboardConnecting The Keyboard Keyboard Description Meter AreaAnnunciators Agilent 60001A MPS Keyboard System Keys  Letter Keys Displaying Status Using The DisplayChanging or Entering Values Sending Scpi Commands Command ExampleQuery Example Error Example Scpi Command Tree Agilent 60001A MPS Keyboard Index AddrProt Unr AmpsLsn Prot Shift SRQ TlkMeas VOLT?, 32 Outp ScpiInit Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates