Agilent Technologies 66001A manual Multiple Loads, Capacitive Loads

Page 25

Figure 2-6. Connecting Power Modules in Series

Each power module has a reverse voltage protection diode across its output. If a reverse voltage is applied, the power module has no control over the current through this diode. To avoid damaging the power module, never connect it to a reverse voltage that can force it to conduct current in excess of the power module’s maximum reverse current (see Table 1-2).

Multiple Loads

When connecting multiple loads to the power module with local sensing, use a separate pair of wires for each load (see Figure 2-7). Each pair of wires should be as short as possible and twisted or bundled to reduce lead inductance and noise pickup. If cabling or terminal considerations require the use of distribution terminals located outside the output connector, then remote voltage sensing is recommended. Connect the sense leads either at the distribution terminals or directly at one of the loads if it is more critical than the others.

Figure 2-7. Connecting Multiple Loads

Optional Relay Connector

If you are using the Option 760 output connector with isolation or polarity reversal relays, the output and sense lead terminal connections are the same as those of the standard output connector. When the output connector Remote/Local switch is in Remote, both the output leads and the sense leads are under relay control. An existing standard connector can be converted into a relay connector. The Installation Guide provided with the connector (see "Related Documents" in Chapter 1) includes instructions for doing this.

Capacitive Loads

Installation 25

Image 25
Contents Agilent Part No Microfiche Part No September USER’S GuideCertification Safety Summary Printing History Symbol DescriptionHerstellerbescheinigung Manufacturers DeclarationUSA Manufacturer’s Name and AddressModel Numbers a 66000A Page Table of Contents Using the Agilent 66001A MPS Keyboard Basic Power Module CommandsOption Description Item DescriptionGeneral Information Accessories Programming Accuracy @ calibration temperature ± 5 C Specifications and Supplemental CharacteristicsOutput Ratings Readback AccuracySupplemental Characteristics Parameter Performance Specifications ParameterAgilent Model 66101A 66102A 66103A 66104A 66105A 66106A Remote Sensing Capability 66101A 66102A Agilent Model 66103A 66104A 66105A 66106ADimensions Recommended Calibration Interval Safety ComplianceWeight Downprogramming Monotonicity Gpib Interface CapabilitiesOutput Characteristic Curve Supplemental CharacteristicsReplaceable Parts List Description Agilent Part No Parameter Agilent Model Number and Parameter ValuePage Switches InstallationInspection Power Module Line Fuse and Switches Switch Setting RI Action Remote Inhibit RI Function SwitchesDisplay Function Switch Calibration Function SwitchesConnecting The Load Installing The Module In The MainframeOutput Connector Output Connections Selecting the Proper Wire SizeResistance Ω/ft AWG No. *AmpacityVoltage Readback Remote Voltage SensingOutput Noise Output RatingParallel Operation Voltlev Outp OFFVoltprot Currlev Voltlev Voltprot Currlev MAX Outp onCapacitive Loads Multiple LoadsBattery Charging Connecting The ControllerInductive Loads OVP ConsiderationsTurn-On Checkout Power Module Front Panel DisplayIntroduction Module Panel DisplayChecking Basic Module Functions Power Module Turn-On StateAddr Dis Prot Error Messages Module Output ConnectorCase Of Trouble Mainframe FailureRuntime Error Messages Selftest Error MessagesMessages DisplaySynopsis Of Commands Basic Power Module CommandsGpib Controller MPS KeyboardPower Module Output Commands Function CommandPower Module Protection Commands Function Power Module Fixed-Mode Output CommandsPower Module Trigger Commands Function Power Module List-Mode Output CommandsControlling List-Mode Output Controlling TriggersUsing the RI/DFI Functions RI Remote Inhibit Input Changing The Power On StateDFI Discrete Fault Indicator Output Command DFI Output Commands FunctionCalibration Changing The Calibration PasswordEnabling Or Disabling The Calibration Function Performing a CalibrationEquipment Required Table A-2. Voltage Calibration ProcedureCalibrating Voltage Keyboard Entry Controller Entry † FunctionDisable the output Table A-3. Current Calibration ProcedureCalcurrlev MIN Calcurrlev MAXRecovering From a Lost Calibration Password Recovering The Factory Calibration ConstantsCalibration Error Messages Agilent Basic Calibration ProgramInitialize Power Supply Assign Power Supply Gpib AddressPassword is Optional only 10 ! Voltage Calibration ProgramPrint Current Calibration not Saved END ifElse Print Current Calibration CompleteConnecting The Keyboard Using The Agilent 66001A MPS KeyboardAgilent 66001A MPS Keyboard Meter Area Keyboard DescriptionAnnunciators Agilent 60001A MPS Keyboard System Keys  Letter Keys Changing or Entering Values Using The DisplayDisplaying Status Command Example Sending Scpi CommandsQuery Example Error Example Scpi Command Tree Agilent 60001A MPS Keyboard Addr IndexProt Unr AmpsShift SRQ Tlk Lsn ProtInit ScpiMeas VOLT?, 32 Outp Agilent Sales and Support Office Agilent Sales and Support OfficesManual Updates