5.In order to maintain proper regulation, the load must always draw at least as much current as is programmed for the module that is in CC mode. You can maintain regulation over a range of current only as long as this condition remains true.

6.As near as is permitted by condition 5, program the current levels of both modules for a balanced distribution of the total output current. Do not program the CV module near its maximum output current level.

7.Program the outputs of both modules ON.

8.When the system is operating, observe that the CV annunciator of the CV module stays on and the CC annunciator on the CC module stays on. If not, repeat steps 1 through 7.

9.Once the modules are properly set up, do all subsequent voltage programming from the CV module; the CC module will track that module.

Remember that although the CV module is controlling the output voltage, the CC module is programmed to a slightly higher voltage. Do not allow the CC module to go into CV mode. Do not

program the CV module to 0 volts without first disabling the outputs of both power modules. Otherwise, the CV module could sink up to 10% of its rated output current from the CC module.

As an example, assume that an Agilent 66102A (20 V @ 7.5 A) and an Agilent 66103A (35 V @ 4.5 A) are connected in parallel to supply a load of 10 amperes at 18 volts. The Agilent 66102A is selected as CC mode module and will supply approximately 7.6 A, maximum. The Agilent 66103A will be the CV module and will supply the remaining current (2.4 A). When the two modules are operated in parallel, you may expect the system to regulate at 18 Volts and 10 amperes as long as the load current remains above 7.6 amperes (see rule 5, above).

The following program shows how to program the two power modules. Generic code (SCPI commands) are shown; they may be sent either from the keyboard or from a GPIB controller.

OUTP OFF

Send to both modules

VOLT:LEV 18

Program the CV mode module

VOLT:PROT 18.5

 

CURR:LEV 2.4

 

VOLT:LEV 19

Program the CC mode module

VOLT:PROT 19.5

 

CURR:LEV MAX

 

OUTP ON

Restore the outputs of both modules

Series Operation

To prevent damage to the equipment, floating voltages must not exceed 240 Vdc. No output terminal may be more than 240 V from chassis ground.

Figure 2-6 shows how the outputs of two power modules may be connected in series to increase the output voltage. Connect only power modules that have the same maximum output current ratings. It is recommended that you set up both modules to operate in CV mode with their current outputs equal to the full load current. If the external load is a storage device, such as a battery or large capacitance, be careful how you shut down the system. For example, turning off just one module could damage the remaining module by subjecting it to double its maximum output voltage from the storage device.

24 Installation

Image 24
Contents USER’S Guide Agilent Part No Microfiche Part No SeptemberCertification Safety Summary Symbol Description Printing HistoryHerstellerbescheinigung Manufacturers DeclarationManufacturer’s Name and Address USAModel Numbers a 66000A Page Table of Contents Basic Power Module Commands Using the Agilent 66001A MPS KeyboardItem Description Option DescriptionGeneral Information Accessories Specifications and Supplemental Characteristics Programming Accuracy @ calibration temperature ± 5 COutput Ratings Readback AccuracyPerformance Specifications Parameter Supplemental Characteristics ParameterAgilent Model 66101A 66102A 66103A 66104A 66105A 66106A 66101A 66102A Agilent Model 66103A 66104A 66105A 66106A Remote Sensing CapabilityRecommended Calibration Interval Safety Compliance DimensionsWeight Downprogramming Monotonicity Gpib Interface CapabilitiesSupplemental Characteristics Output Characteristic CurveReplaceable Parts List Description Agilent Part No Parameter Agilent Model Number and Parameter ValuePage Installation SwitchesInspection Power Module Line Fuse and Switches Remote Inhibit RI Function Switches Switch Setting RI ActionDisplay Function Switch Calibration Function SwitchesInstalling The Module In The Mainframe Connecting The LoadOutput Connector Selecting the Proper Wire Size Output ConnectionsResistance Ω/ft AWG No. *AmpacityRemote Voltage Sensing Voltage ReadbackOutput Rating Output NoiseParallel Operation Outp OFF VoltlevVoltprot Currlev Voltlev Voltprot Currlev MAX Outp onMultiple Loads Capacitive LoadsConnecting The Controller Battery ChargingInductive Loads OVP ConsiderationsPower Module Front Panel Display Turn-On CheckoutIntroduction Module Panel DisplayPower Module Turn-On State Checking Basic Module FunctionsAddr Dis Prot Module Output Connector Error MessagesCase Of Trouble Mainframe FailureSelftest Error Messages Runtime Error MessagesMessages DisplayBasic Power Module Commands Synopsis Of CommandsGpib Controller MPS KeyboardCommand Power Module Output Commands FunctionPower Module Protection Commands Function Power Module Fixed-Mode Output CommandsPower Module List-Mode Output Commands Power Module Trigger Commands FunctionControlling List-Mode Output Controlling TriggersChanging The Power On State Using the RI/DFI Functions RI Remote Inhibit InputDFI Discrete Fault Indicator Output Command DFI Output Commands FunctionChanging The Calibration Password CalibrationEnabling Or Disabling The Calibration Function Performing a CalibrationTable A-2. Voltage Calibration Procedure Equipment RequiredCalibrating Voltage Keyboard Entry Controller Entry † FunctionTable A-3. Current Calibration Procedure Disable the outputCalcurrlev MIN Calcurrlev MAXRecovering The Factory Calibration Constants Recovering From a Lost Calibration PasswordCalibration Error Messages Agilent Basic Calibration ProgramAssign Power Supply Gpib Address Initialize Power SupplyPassword is Optional only 10 ! Voltage Calibration ProgramEND if Print Current Calibration not SavedElse Print Current Calibration CompleteUsing The Agilent 66001A MPS Keyboard Connecting The KeyboardAgilent 66001A MPS Keyboard Keyboard Description Meter AreaAnnunciators Agilent 60001A MPS Keyboard System Keys  Letter Keys Using The Display Changing or Entering ValuesDisplaying Status Sending Scpi Commands Command ExampleQuery Example Error Example Scpi Command Tree Agilent 60001A MPS Keyboard Index AddrProt Unr AmpsLsn Prot Shift SRQ TlkScpi InitMeas VOLT?, 32 Outp Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates