ENTRY KEYS

These keys let you enter numeric values. Values can be entered directly. If the Meter function is active, the display automatically switches to the Command function with the appropriate command for programming the metered mode on the display.

This key enters the decimal point.

This key is used to separate multiple command arguments such as those used with a List command.

These keys let you increment and decrement the numeric values in the selected field when the Meter function is active. Note that numeric values change in larger steps the longer these keys are held down. When the Command function is active, these keys let you increment and decrement numeric command parameters.

These keys let you assign a polarity to a value that is entered with the numeric Entry keys. You must enter the polarity before you enter the number.

This key lets you enter the exponent of a value that is entered with the numeric Entry keys. Enter the exponent after you enter the number.

This key backspaces over and deletes individual digits in a numeric entry field (same as ).

This key clears the entire numeric entry field.

This key executes a command.

LETTER KEYS

These keys let you select SCPI commands when the Command function is active. Letter keys select the choices that begin with the letter that is pressed and move across the command tree when

additional selections are available. ed letters make selections by scrolling through the choices at the present command level that start with the shifted letter.

Note

Nothing happens if a letter is pressed that has no corresponding SCPI command.

This key is the prompt for the SCPI instrument commands.

This key is the prompt for the GPIB commands (IEEE-488.2).

This key is the prompt for SCPI queries. Append the prompt to the query. The result of the query appears on the display after it is entered.

This key accesses the shifted key functions. The Shift annunciator on the display indicates that the key was depressed. The annunciator turns off when the next key is pressed.

This key locks the shift key functions. The Shift annunciator on the display indicates when the shift functions are locked.

This key backspaces over and deletes individual digits in a numeric entry field (same as ).

This key executes a command.

44 Agilent 60001A MPS Keyboard

Image 44
Contents USER’S Guide Agilent Part No Microfiche Part No SeptemberCertification Safety Summary Symbol Description Printing HistoryHerstellerbescheinigung Manufacturers DeclarationModel Numbers a 66000A Manufacturer’s Name and AddressUSA Page Table of Contents Basic Power Module Commands Using the Agilent 66001A MPS KeyboardGeneral Information Item DescriptionOption Description Accessories Specifications and Supplemental Characteristics Programming Accuracy @ calibration temperature ± 5 COutput Ratings Readback AccuracyAgilent Model 66101A 66102A 66103A 66104A 66105A 66106A Performance Specifications ParameterSupplemental Characteristics Parameter 66101A 66102A Agilent Model 66103A 66104A 66105A 66106A Remote Sensing CapabilityRecommended Calibration Interval Safety Compliance DimensionsWeight Downprogramming Monotonicity Gpib Interface CapabilitiesSupplemental Characteristics Output Characteristic CurveReplaceable Parts List Description Agilent Part No Parameter Agilent Model Number and Parameter ValuePage Inspection InstallationSwitches Power Module Line Fuse and Switches Remote Inhibit RI Function Switches Switch Setting RI ActionDisplay Function Switch Calibration Function SwitchesOutput Connector Installing The Module In The MainframeConnecting The Load Selecting the Proper Wire Size Output ConnectionsResistance Ω/ft AWG No. *AmpacityRemote Voltage Sensing Voltage ReadbackParallel Operation Output RatingOutput Noise Outp OFF VoltlevVoltprot Currlev Voltlev Voltprot Currlev MAX Outp onMultiple Loads Capacitive LoadsConnecting The Controller Battery ChargingInductive Loads OVP ConsiderationsPower Module Front Panel Display Turn-On CheckoutIntroduction Module Panel DisplayAddr Dis Prot Power Module Turn-On StateChecking Basic Module Functions Module Output Connector Error MessagesCase Of Trouble Mainframe FailureSelftest Error Messages Runtime Error MessagesMessages DisplayBasic Power Module Commands Synopsis Of CommandsGpib Controller MPS KeyboardCommand Power Module Output Commands FunctionPower Module Protection Commands Function Power Module Fixed-Mode Output CommandsPower Module List-Mode Output Commands Power Module Trigger Commands FunctionControlling List-Mode Output Controlling TriggersChanging The Power On State Using the RI/DFI Functions RI Remote Inhibit InputDFI Discrete Fault Indicator Output Command DFI Output Commands FunctionChanging The Calibration Password CalibrationEnabling Or Disabling The Calibration Function Performing a CalibrationTable A-2. Voltage Calibration Procedure Equipment RequiredCalibrating Voltage Keyboard Entry Controller Entry † FunctionTable A-3. Current Calibration Procedure Disable the outputCalcurrlev MIN Calcurrlev MAXRecovering The Factory Calibration Constants Recovering From a Lost Calibration PasswordCalibration Error Messages Agilent Basic Calibration ProgramAssign Power Supply Gpib Address Initialize Power SupplyPassword is Optional only 10 ! Voltage Calibration ProgramEND if Print Current Calibration not SavedElse Print Current Calibration CompleteAgilent 66001A MPS Keyboard Using The Agilent 66001A MPS KeyboardConnecting The Keyboard Keyboard Description Meter AreaAnnunciators Agilent 60001A MPS Keyboard System Keys  Letter Keys Displaying Status Using The DisplayChanging or Entering Values Sending Scpi Commands Command ExampleQuery Example Error Example Scpi Command Tree Agilent 60001A MPS Keyboard Index AddrProt Unr AmpsLsn Prot Shift SRQ TlkMeas VOLT?, 32 Outp ScpiInit Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates