Agilent Technologies 66001A 10 ! Voltage Calibration Program, Assign Power Supply Gpib Address

Page 39

10 ! VOLTAGE CALIBRATION PROGRAM

20 ! CONNECT INSTRUMENTS AS SHOWN IN FIGURE A-la

30 !

40DIM Resp$[255],Err_msg$[255]

50Volt_cal: !

60Err_found=0

70INPUT "CONNECT INSTRUMENTS AS SHOWN IN FIGURE A-la ... Press any key to continue.” ,Resp$

80Resp$=""

90

ASSIGN @Ps TO 70500

! ASSIGN POWER SUPPLY GPIB ADDRESS

100

 

! NOTE: DEFAULT MODULE SLOT IS 0

110

 

!

120

OUTPUT @Ps;"*RST;OUTPUT ON"

! INITIALIZE POWER SUPPLY

130OUTPUT @Ps;"VOLT MAX"

140OUTPUT @Ps;"CURR MAX"

150

OUTPUT @Ps;"CAL:STATE ON, 66101"

! PASSWORD IS OPTIONAL - ONLY

160

 

! REQUIRED IF SET TO NON-ZERO

170

 

! NOTE: DEFAULT = MODEL NUMBER

180

OUTPUT @Ps;"CAL:VOLTAGE:LEVEL MIN"

 

190INPUT "ENTER VOLTAGE MEASUREMENT FROM EXTERNAL VOLTMETER",Volt_read

200OUTPUT @Ps;"CAL:VOLTAGE ";Volt_read

210OUTPUT @Ps;"CAL:VOLTAGE:LEVEL MAX"

220INPUT "ENTER VOLTAGE MEASUREMENT FROM EXTERNAL VOLTMETER",Volt_read

230OUTPUT @Ps;"CAL:VOLTAGE ";Volt_read

240

OUTPUT @Ps;"CAL:VOLTAGE:PROT"

! CALIBRATE OVERVOLTAGE PROTECTION

250

 

! CIRCUITRY

260GOSUB Save_cal

270IF Err_found THEN

280INPUT "ERRORS have occurred, REPEAT VOLTAGE CALIBRATION (Y OR N)?",Resp$

290IF TRIM$(UPC$(Resp$[ l,l ]))="Y" THEN GOTO Volt_cal

300END IF

310IF Err_found THEN

320PRINT "VOLTAGE CALIBRATION NOT SAVED"

330ELSE

340PRINT "VOLTAGE CALIBRATION COMPLETE”

350END

360!

370! CURRENT CALIBRATION PROGRAM

380! CONNECT INSTRUMENTS AS SHOWN IN FIGURE A-lb

390!

400Current_cal: !

410Err_found=0

420INPUT "CONNECT INSTRUMENTS AS SHOWN IN FIGURE A-lb ... Press any key to continue.",Resp$

430Resp$=""

440

OUTPUT @Ps;"CAL:STATE ON, 66101"

! PASSWORD IS OPTIONAL - ONLY

450

 

! REQUIRED IF SET TO NON-ZERO

460

 

! NOTE: DEFAULT = MODEL NUMBER

470

 

!

480! REFER TO TABLE A-1 FOR CORRECT SHUNT VALUE FOR MODEL BEING CALIBRATED

490 !

Figure A-2. Agilent BASIC Calibration Program

Calibration 39

Image 39
Contents Agilent Part No Microfiche Part No September USER’S GuideCertification Safety Summary Manufacturers Declaration Symbol DescriptionPrinting History HerstellerbescheinigungManufacturer’s Name and Address USAModel Numbers a 66000A Page Table of Contents Using the Agilent 66001A MPS Keyboard Basic Power Module CommandsItem Description Option DescriptionGeneral Information Accessories Readback Accuracy Specifications and Supplemental CharacteristicsProgramming Accuracy @ calibration temperature ± 5 C Output RatingsPerformance Specifications Parameter Supplemental Characteristics ParameterAgilent Model 66101A 66102A 66103A 66104A 66105A 66106A Remote Sensing Capability 66101A 66102A Agilent Model 66103A 66104A 66105A 66106ADownprogramming Monotonicity Gpib Interface Capabilities Recommended Calibration Interval Safety ComplianceDimensions WeightParameter Agilent Model Number and Parameter Value Supplemental CharacteristicsOutput Characteristic Curve Replaceable Parts List Description Agilent Part NoPage Installation SwitchesInspection Power Module Line Fuse and Switches Calibration Function Switches Remote Inhibit RI Function SwitchesSwitch Setting RI Action Display Function SwitchInstalling The Module In The Mainframe Connecting The LoadOutput Connector AWG No. *Ampacity Selecting the Proper Wire SizeOutput Connections Resistance Ω/ftVoltage Readback Remote Voltage SensingOutput Rating Output NoiseParallel Operation Voltprot Currlev MAX Outp on Outp OFFVoltlev Voltprot Currlev VoltlevCapacitive Loads Multiple LoadsOVP Considerations Connecting The ControllerBattery Charging Inductive LoadsModule Panel Display Power Module Front Panel DisplayTurn-On Checkout IntroductionPower Module Turn-On State Checking Basic Module FunctionsAddr Dis Prot Mainframe Failure Module Output ConnectorError Messages Case Of TroubleDisplay Selftest Error MessagesRuntime Error Messages MessagesMPS Keyboard Basic Power Module CommandsSynopsis Of Commands Gpib ControllerPower Module Fixed-Mode Output Commands CommandPower Module Output Commands Function Power Module Protection Commands FunctionControlling Triggers Power Module List-Mode Output CommandsPower Module Trigger Commands Function Controlling List-Mode OutputDFI Output Commands Function Changing The Power On StateUsing the RI/DFI Functions RI Remote Inhibit Input DFI Discrete Fault Indicator Output CommandPerforming a Calibration Changing The Calibration PasswordCalibration Enabling Or Disabling The Calibration FunctionKeyboard Entry Controller Entry † Function Table A-2. Voltage Calibration ProcedureEquipment Required Calibrating VoltageCalcurrlev MAX Table A-3. Current Calibration ProcedureDisable the output Calcurrlev MINAgilent Basic Calibration Program Recovering The Factory Calibration ConstantsRecovering From a Lost Calibration Password Calibration Error Messages10 ! Voltage Calibration Program Assign Power Supply Gpib AddressInitialize Power Supply Password is Optional onlyPrint Current Calibration Complete END ifPrint Current Calibration not Saved ElseUsing The Agilent 66001A MPS Keyboard Connecting The KeyboardAgilent 66001A MPS Keyboard Agilent 60001A MPS Keyboard Keyboard DescriptionMeter Area Annunciators System Keys  Letter Keys Using The Display Changing or Entering ValuesDisplaying Status Command Example Sending Scpi CommandsQuery Example Error Example Scpi Command Tree Agilent 60001A MPS Keyboard Amps IndexAddr Prot UnrShift SRQ Tlk Lsn ProtScpi InitMeas VOLT?, 32 Outp Agilent Sales and Support Office Agilent Sales and Support OfficesManual Updates