Agilent Technologies 66001A Controlling the Output State, Command, Controlling Fixed-Mode Output

Page 32

Controlling the Output State

The following commands control the power module output state:

Command

OUTP ON

OUTP OFF

OUTP1,NOR

OUTP 0,NOR

OUTP: REL 1

OUTP:REL 0 OUTP:REL:POL NORM OUTP:REL:POL REV

Table 4-1. Power Module Output Commands

Function

Enables the output. Disables the output.

(If an optional output relay is installed, the command is executed in a sequence that prevents "hot switching".)

Enables the output without affecting the output relay. Disables the output without affecting the output relay. Closes the (optional) output relay contacts.

Opens the (optional) output relay contacts.

Makes the (optional) output relay polarity the same as the module output.

Makes the (optional) output relay polarity opposite to the module output. The output is disabled while the relay polarity is being reversed.)

Controlling Protection Functions

The following commands control the power module protection circuits.

Command

CURR:PROT:STAT ON

CURR: PROT: STAT OFF VOLT:PROT <value>

OUTP:PROT:CLE

OUTP:PROT:DEL .05

Table 4-2. Power Module Protection Commands

Function

Turns on the OCP (overcurrent protection), which disables the output if the output current exceeds the programmed current.

Turns off the current protection.

Programs the OVP (overvoltage protection) level. The protection circuit will trip if the output voltage exceeds the OVP level.

Clears any active protection conditions (panel Prot annunciator goes off). Before the command can comply, you must remove the condition that caused the protection circuit to trip.

Specifies a delay (in seconds) between the time a protection condition is detected and the actual activation of the protection circuit. (Does not apply to OVP.)

Controlling Fixed-Mode Output

The following commands control the power module fixed-mode voltage and current output.

 

Table 4-3. Power Module Fixed-Mode Output Commands

Command

 

Function

CURR:MODE FIX

CURR <value> CURR:TRIG <value> MEAS:CURR?

VOLT:MODE FIX

VOLT <value> VOLT: TRIG <value> MEAS:VOLT? VOLT:SENS?

Sets the current mode to fixed (as opposed to list). FIX is the factory default power on state.

Programs the immediate output current level (in amperes).

Programs the triggered output current level (in amperes). See "Trigger Commands". Returns the present value of the output current.

Sets the voltage mode to fixed (as opposed to list). FIX is the factory default power on state.

Programs the immediate output voltage level (in amperes).

Programs the triggered output voltage level (in amperes). See "Trigger Commands". Returns the present value of the output voltage.

Returns the position of the output connector sense switch (INT is local; EXT is remote).

32 Basic Power Module Commands

Image 32
Contents USER’S Guide Agilent Part No Microfiche Part No SeptemberCertification Safety Summary Symbol Description Printing HistoryHerstellerbescheinigung Manufacturers DeclarationModel Numbers a 66000A Manufacturer’s Name and AddressUSA Page Table of Contents Basic Power Module Commands Using the Agilent 66001A MPS KeyboardGeneral Information Item DescriptionOption Description Accessories Specifications and Supplemental Characteristics Programming Accuracy @ calibration temperature ± 5 COutput Ratings Readback AccuracyAgilent Model 66101A 66102A 66103A 66104A 66105A 66106A Performance Specifications ParameterSupplemental Characteristics Parameter 66101A 66102A Agilent Model 66103A 66104A 66105A 66106A Remote Sensing CapabilityRecommended Calibration Interval Safety Compliance DimensionsWeight Downprogramming Monotonicity Gpib Interface CapabilitiesSupplemental Characteristics Output Characteristic CurveReplaceable Parts List Description Agilent Part No Parameter Agilent Model Number and Parameter ValuePage Inspection InstallationSwitches Power Module Line Fuse and Switches Remote Inhibit RI Function Switches Switch Setting RI ActionDisplay Function Switch Calibration Function SwitchesOutput Connector Installing The Module In The MainframeConnecting The Load Selecting the Proper Wire Size Output ConnectionsResistance Ω/ft AWG No. *AmpacityRemote Voltage Sensing Voltage ReadbackParallel Operation Output RatingOutput Noise Outp OFF VoltlevVoltprot Currlev Voltlev Voltprot Currlev MAX Outp onMultiple Loads Capacitive LoadsConnecting The Controller Battery ChargingInductive Loads OVP ConsiderationsPower Module Front Panel Display Turn-On CheckoutIntroduction Module Panel DisplayAddr Dis Prot Power Module Turn-On StateChecking Basic Module Functions Module Output Connector Error MessagesCase Of Trouble Mainframe FailureSelftest Error Messages Runtime Error MessagesMessages DisplayBasic Power Module Commands Synopsis Of CommandsGpib Controller MPS KeyboardCommand Power Module Output Commands FunctionPower Module Protection Commands Function Power Module Fixed-Mode Output CommandsPower Module List-Mode Output Commands Power Module Trigger Commands FunctionControlling List-Mode Output Controlling TriggersChanging The Power On State Using the RI/DFI Functions RI Remote Inhibit InputDFI Discrete Fault Indicator Output Command DFI Output Commands FunctionChanging The Calibration Password CalibrationEnabling Or Disabling The Calibration Function Performing a CalibrationTable A-2. Voltage Calibration Procedure Equipment RequiredCalibrating Voltage Keyboard Entry Controller Entry † FunctionTable A-3. Current Calibration Procedure Disable the outputCalcurrlev MIN Calcurrlev MAXRecovering The Factory Calibration Constants Recovering From a Lost Calibration PasswordCalibration Error Messages Agilent Basic Calibration ProgramAssign Power Supply Gpib Address Initialize Power SupplyPassword is Optional only 10 ! Voltage Calibration ProgramEND if Print Current Calibration not SavedElse Print Current Calibration CompleteAgilent 66001A MPS Keyboard Using The Agilent 66001A MPS KeyboardConnecting The Keyboard Keyboard Description Meter AreaAnnunciators Agilent 60001A MPS Keyboard System Keys  Letter Keys Displaying Status Using The DisplayChanging or Entering Values Sending Scpi Commands Command ExampleQuery Example Error Example Scpi Command Tree Agilent 60001A MPS Keyboard Index AddrProt Unr AmpsLsn Prot Shift SRQ TlkMeas VOLT?, 32 Outp ScpiInit Agilent Sales and Support Offices Agilent Sales and Support OfficeManual Updates