Agilent Technologies 66001A manual  System Keys

Page 43

([Shift] [Meter])

([Shift] [Command])

([Shift] [Status])

SYSTEM KEYS

This key switches from remote operation (over the GPIB) to local operation using the keyboard.

Press when the Rmt annunciator is on to enable the keyboard.

This key moves the cursor to the subaddress field so you can select another module. Use the Entry or Select keys to enter another subaddress.

FUNCTION KEYS

This function continuously displays operating information about the selected module. The Meter function is the turn-on default function, and is also in effect whenever the mainframe is in remote mode.

This function lets you send a SCPI command or query to the selected module. Use the letter keys or the Select keys to select the SCPI command. Refer to the power module Programming Guide for a complete description of the SCPI commands. Refer to the end of this appendix for the SCPI command tree used by the keyboard.

This function displays status information about the following QUEStionable and CONDition status register bits: CV, CC, OV, OC, OT, RT, WTG, and DWE. Refer to the power module Programming Guide for a complete description of the QUEStionable and CONDition status registers.

This function displays the error messages whenever the Err annunciator on the display indicates that an error has occurred.

This function lets you format the meter area of the display. Power supply modules do not use additional meter formats.

This function lets you send the next command entered from the keypad to all the modules in the mainframe. The subaddress field on the display indicates "ALL" when this function is active. You

must press each time you want to send a command to all modules.

This function clears any latched protection features that have tripped.

SELECT KEYS

When the Meter function is active, these keys let you move across the display so that you can enter or change values in the selected fields. When the Command function is active, these keys let you move across the various levels of the command tree structure. Numeric command parameters default either to zero or the last programmed value.

When the Meter function is active, these keys let you increment and decrement the numeric values in the selected field. When the Command function is active, these keys let you scroll up and down through the choices at each level of the command tree.

This key lets you recall up to five previous commands that were sent. You can edit any recalled

command. Press to send the command.

This key exits any action and returns the display to the Meter function. If the Command function is

active, the displayed command is not executed (same as ).

Agilent 66001A MPS Keyboard 43

Image 43
Contents Agilent Part No Microfiche Part No September USER’S GuideCertification Safety Summary Manufacturers Declaration Symbol DescriptionPrinting History HerstellerbescheinigungUSA Manufacturer’s Name and AddressModel Numbers a 66000A Page Table of Contents Using the Agilent 66001A MPS Keyboard Basic Power Module CommandsOption Description Item DescriptionGeneral Information Accessories Readback Accuracy Specifications and Supplemental CharacteristicsProgramming Accuracy @ calibration temperature ± 5 C Output RatingsSupplemental Characteristics Parameter Performance Specifications ParameterAgilent Model 66101A 66102A 66103A 66104A 66105A 66106A Remote Sensing Capability 66101A 66102A Agilent Model 66103A 66104A 66105A 66106ADownprogramming Monotonicity Gpib Interface Capabilities Recommended Calibration Interval Safety ComplianceDimensions WeightParameter Agilent Model Number and Parameter Value Supplemental CharacteristicsOutput Characteristic Curve Replaceable Parts List Description Agilent Part NoPage Switches InstallationInspection Power Module Line Fuse and Switches Calibration Function Switches Remote Inhibit RI Function SwitchesSwitch Setting RI Action Display Function SwitchConnecting The Load Installing The Module In The MainframeOutput Connector AWG No. *Ampacity Selecting the Proper Wire SizeOutput Connections Resistance Ω/ftVoltage Readback Remote Voltage SensingOutput Noise Output RatingParallel Operation Voltprot Currlev MAX Outp on Outp OFFVoltlev Voltprot Currlev VoltlevCapacitive Loads Multiple LoadsOVP Considerations Connecting The ControllerBattery Charging Inductive LoadsModule Panel Display Power Module Front Panel DisplayTurn-On Checkout IntroductionChecking Basic Module Functions Power Module Turn-On StateAddr Dis Prot Mainframe Failure Module Output ConnectorError Messages Case Of TroubleDisplay Selftest Error MessagesRuntime Error Messages MessagesMPS Keyboard Basic Power Module CommandsSynopsis Of Commands Gpib ControllerPower Module Fixed-Mode Output Commands CommandPower Module Output Commands Function Power Module Protection Commands FunctionControlling Triggers Power Module List-Mode Output CommandsPower Module Trigger Commands Function Controlling List-Mode OutputDFI Output Commands Function Changing The Power On StateUsing the RI/DFI Functions RI Remote Inhibit Input DFI Discrete Fault Indicator Output CommandPerforming a Calibration Changing The Calibration PasswordCalibration Enabling Or Disabling The Calibration FunctionKeyboard Entry Controller Entry † Function Table A-2. Voltage Calibration ProcedureEquipment Required Calibrating VoltageCalcurrlev MAX Table A-3. Current Calibration ProcedureDisable the output Calcurrlev MINAgilent Basic Calibration Program Recovering The Factory Calibration ConstantsRecovering From a Lost Calibration Password Calibration Error Messages10 ! Voltage Calibration Program Assign Power Supply Gpib AddressInitialize Power Supply Password is Optional onlyPrint Current Calibration Complete END ifPrint Current Calibration not Saved ElseConnecting The Keyboard Using The Agilent 66001A MPS KeyboardAgilent 66001A MPS Keyboard Agilent 60001A MPS Keyboard Keyboard DescriptionMeter Area Annunciators System Keys  Letter Keys Changing or Entering Values Using The DisplayDisplaying Status Command Example Sending Scpi CommandsQuery Example Error Example Scpi Command Tree Agilent 60001A MPS Keyboard Amps IndexAddr Prot UnrShift SRQ Tlk Lsn ProtInit ScpiMeas VOLT?, 32 Outp Agilent Sales and Support Office Agilent Sales and Support OfficesManual Updates