3654 Portable H2/N2 Analyzer - Operating Instructions

25 of 66

 

 

2.5 Monitoring Measurements In Real-Time

You may wish to analyze a particular sampling point via the WIN3654 program's Monitoring menu. To use this Monitoring chart, the analyzer must be connected to your PC.

Choose Monitoring from the WIN3654 menu to bring up a chart display like the one illustrated in Fig 2-4.

Fig 2-4:Real-Time Monitoring

The Monitoring chart shows the gas concentration (in blue), temperature (in red), and pressure (in green) as the sample is being measured by the 3654 instrument. The chart is updated directly from instrument measurements, at a rate determined by the time scale set in the TIMEBASE box at the lower right corner of the chart.

Click the TIMEBASE up/down pointers to change the time scale of the divisions of the chart. Each division mark along the baseline (1, 2, ...10) can be made to represent from 30 seconds to 2½ hours, providing from 5 minutes to 25 hours of continuously displayed samples. The chart updating rate is determined by the time scale selected, as shown in Table 2-1, “Chart Updating Rate,” on page 25.

Table 2-1:Chart Updating Rate

 

 

Timebase

Updating Rate*

Maximum Samples (10 divisions)

 

 

30

Seconds/Division

5 Seconds/Sample

60

 

 

 

 

 

 

 

1 Minute/Division

5 Seconds/Sample

120

 

 

 

 

 

 

 

 

10

Minutes/Division

5 Seconds/Sample

1,200

 

 

 

 

 

 

 

 

30

Minutes/Division

9 Seconds/Sample

2,000

 

 

 

 

 

 

 

1 Hour/Division

18 Seconds/Sample

2,000

 

 

 

 

 

 

 

2.5 Hours/Division

45 Seconds/Sample

2,000

 

 

 

 

 

 

 

*This chart's updating rate is independent from the acquisition rate (see

 

 

“Automatic Data Acquisition - Setting Sampling Intervals” on page 31).

 

 

 

 

 

 

 

 

 

 

 

 

 

Operator Manual

 

 

ORBISPHERE

Image 31
Contents Operator Manual Page Product Recycling Information Orbisphere Following only applies to European customers Product DisposalRestriction of Hazardous Substances Board Analog Board Spacer External ConnectorTable of Contents Maintenance and Troubleshooting Accessories and AttachmentsCalibrations Part Lists SpecificationsAppendix a Glossary Table of Contents 3654 Portable H2/N2 Analyzer Revision History Safety ConventionsManual Overview About this ManualSafety Recommendations Safety PrecautionsService and Repairs Acknowledgements Precautionary LabelsOrbisphere What you have Received InstallationSensor Installation WIN3654 PC Program InstallationAtmospheric Pressure Equilibrium Instrument PC Connection External Power Supply optionalPurge Gas Connection ConnectionsPin RS232 Signal Use Flow Chamber Installation Installation Completion Check List Dissolved Gas Measurements Flow ChamberOrbisphere Operating Controls Operating InstructionsMeasurement Phase on Measurement Mode Select Gas Measurement PhaseTaking Measurements Automatic Data Acquisition Storing Measurements In the InstrumentManual Data Acquisition Viewing Stored Measurements Storing and Accessing Measurements From a ComputerDownloading Stored Values Copying Values Altering the Sampling Point DescriptionsSaving Values Clearing Stored Values Printing ValuesTimebase Updating Rate Maximum Samples 10 divisions Monitoring Measurements In Real-TimeAfter Use and Storage Main Menu Basics Options SetupOptions Setup 3654 Portable H2/N2 Analyzer Analyzer PC Connection Reviewing Instrument ConfigurationConfiguring the Instrument Gas Measurement PhaseMeasurement Units Membrane Selection Automatic Data Acquisition Setting Sampling IntervalsLocking Out the Instrument’s CAL Button Calibration MediumRolling Average Automatic Shutdown Orbisphere Barometric Pressure Sensor Calibration CalibrationsCalibration Method Time Benefits Drawbacks Reference Sensor CalibrationCalibration Setup Calibration In H2 or N2 at Atmospheric Pressure Calibration In H2 or N2 at Elevated Pressure Calibrating In a Liquid H2 or N2 Solution Model 32311 Flow Meter Accessories and AttachmentsModel 32939 External Power Supply Maintenance Schedule Maintenance and TroubleshootingPower Supply Sensor MaintenanceMounting Dual Membranes Membrane ReplacementRemoving the Membrane Installing the Membrane Orbisphere Place the membrane holding ring on the installation tool tip 1 N2 Purge Gas Purge Gas Cylinder RefillFilling the Purge Gas Cylinder 2 CO2 Purge GasSerial Test Troubleshooting InstrumentKeyboard Test Display Test Clock SettingsMeasurements View Analog Voltages ViewContinuous Purge Troubleshooting OperationDiagnostics Messages Program IdentificationSymptom Cause Possible Solution Troubleshooting TableInstrument Specifications SpecificationsSensor Type 31 270 H Sensor SpecificationsPFA Etfe Ectfe Sensor Type 31 272 H 31 570 NSpecifications 3654 Portable H2/N2 Analyzer Analyzer Spare Parts Instrument ConfigurationsTC Sensor and Parts on Configured System Part ListsSensor Spare Parts Unit Appendix a GlossaryCommon Units Terms and DefinitionsGlossary 3654 Portable H2/N2 Analyzer Tables and Illustrations AnnexAnnex 3654 Portable H2/N2 Analyzer User Notes