3654 Portable H2/N2 Analyzer - Maintenance and Troubleshooting | 53 of 66 |
|
|
6.6Troubleshooting - Operation
6.6.1Program Identification
In case of problems, it may be necessary to check the current version of your software as this may be required by Hach Ultra if you need to contact them for assistance in problem solving.
To identify the version of the software in your instrument, switch off the instrument. Then hold down the MEAS button while switching the instrument back on. The instrument will then briefly display the software version number before going through it’s
6.6.2 Diagnostics Messages
The following are warning messages that may appear on the instrument LCD in place of the gas concentration.
Table
Message | Meaning | |
| This message appears on the LCD if the purge gas supply fails. To detect this | |
Pur | condition, the voltage signal from the TC sensor is measured. This is normally | |
between | ||
| If the flow of purge gas stops, the sensor voltage wanders outside the permitted | |
| range of 100mV to 4 V. See also “Continuous Purge” below. | |
|
| |
Out | This message appears if the sensor is unplugged from its correct position, or if the | |
temperature is outside the range | ||
|
| |
| During | |
| tests successfully (see “Taking Measurements” on page 19). If the problem per- | |
Err | sists, contact your local Hach Ultra service representative for assistance. | |
| ||
During calibration, this message indicates that the partial pressure of gas is below | ||
| ||
| the acceptable range (<5 mbar), or that you have pressed a wrong key, as | |
| explained in “Sensor Calibration” on page 36. | |
|
|
6.6.3 Continuous Purge
It may be useful, particularly when seeing a [Pur] error message on your display, to view the purge voltage of the TC sensor in a continuous purge mode. In this mode, there is no purge/measurement cycle, as the purge gas is flowing through the sensor continuously.
To start continuous purge mode, put the instrument into standby, then hold down the Down Arrow button while switching the instrument back on again. The instrument displays [Pur], followed by the purge gas voltage.
At
To exit from continuous purge mode, put the instrument in standby, then switch back on again without holding any keys.
Operator Manual | ORBISPHERE |