Hach 3654 Portable H2/N2 Analyzer manual Calibrations, Barometric Pressure Sensor Calibration

Page 41

3654 Portable H2/N2 Analyzer - Calibrations

35 of 66

 

 

4Calibrations

4.1 Barometric Pressure Sensor Calibration

You must have access to an accurate barometer to calibrate the instrument's internal barometric pressure sensor. This is done using the PC WIN3654 program.

Choose Troubleshooting, Pressure Calibration and a message will warn that the current calibration will be lost.

Choose OK to continue.

The calibration procedure then displays a Pressure Calibration dialog box (as illustrated in Fig 4-1). The Measured Pressure value shows the current instrument pressure reading.

Fig 4-1:Pressure Calibration Screens 2 & 3

Enter the current atmospheric pressure, in mbars, in the Calibration Pressure entry box. Choose Calibrate to direct the instrument to read and display the Measured Pressure using this calibration value.

Choose Quit when you are satisfied with the pressure calibration to return to normal operation.

Operator Manual

ORBISPHERE

Image 41
Contents Operator Manual Page Product Recycling Information Orbisphere Following only applies to European customers Product DisposalRestriction of Hazardous Substances Board Analog Board Spacer External ConnectorTable of Contents Calibrations Accessories and AttachmentsMaintenance and Troubleshooting Appendix a Glossary SpecificationsPart Lists Table of Contents 3654 Portable H2/N2 Analyzer Manual Overview Safety ConventionsAbout this Manual Revision HistoryService and Repairs Safety PrecautionsSafety Recommendations Acknowledgements Precautionary LabelsOrbisphere What you have Received InstallationAtmospheric Pressure Equilibrium WIN3654 PC Program InstallationSensor Installation Purge Gas Connection External Power Supply optionalConnections Instrument PC ConnectionPin RS232 Signal Use Flow Chamber Installation Installation Completion Check List Dissolved Gas Measurements Flow ChamberOrbisphere Operating Controls Operating InstructionsMeasurement Phase on Taking Measurements Select Gas Measurement PhaseMeasurement Mode Automatic Data Acquisition Storing Measurements In the InstrumentManual Data Acquisition Downloading Stored Values Storing and Accessing Measurements From a ComputerViewing Stored Measurements Saving Values Altering the Sampling Point DescriptionsCopying Values Clearing Stored Values Printing ValuesTimebase Updating Rate Maximum Samples 10 divisions Monitoring Measurements In Real-TimeAfter Use and Storage Main Menu Basics Options SetupOptions Setup 3654 Portable H2/N2 Analyzer Configuring the Instrument Reviewing Instrument ConfigurationGas Measurement Phase Analyzer PC ConnectionMeasurement Units Membrane Selection Automatic Data Acquisition Setting Sampling IntervalsRolling Average Calibration MediumLocking Out the Instrument’s CAL Button Automatic Shutdown Orbisphere Barometric Pressure Sensor Calibration CalibrationsCalibration Method Time Benefits Drawbacks Reference Sensor CalibrationCalibration Setup Calibration In H2 or N2 at Atmospheric Pressure Calibration In H2 or N2 at Elevated Pressure Calibrating In a Liquid H2 or N2 Solution Model 32311 Flow Meter Accessories and AttachmentsModel 32939 External Power Supply Power Supply Maintenance and TroubleshootingSensor Maintenance Maintenance ScheduleRemoving the Membrane Membrane ReplacementMounting Dual Membranes Installing the Membrane Orbisphere Place the membrane holding ring on the installation tool tip 1 N2 Purge Gas Purge Gas Cylinder RefillFilling the Purge Gas Cylinder 2 CO2 Purge GasKeyboard Test Troubleshooting InstrumentSerial Test Display Test Clock SettingsMeasurements View Analog Voltages ViewDiagnostics Messages Troubleshooting OperationProgram Identification Continuous PurgeSymptom Cause Possible Solution Troubleshooting TableInstrument Specifications SpecificationsSensor Type 31 270 H Sensor SpecificationsPFA Etfe Ectfe Sensor Type 31 272 H 31 570 NSpecifications 3654 Portable H2/N2 Analyzer TC Sensor and Parts on Configured System Instrument ConfigurationsPart Lists Analyzer Spare PartsSensor Spare Parts Common Units Appendix a GlossaryTerms and Definitions UnitGlossary 3654 Portable H2/N2 Analyzer Tables and Illustrations AnnexAnnex 3654 Portable H2/N2 Analyzer User Notes