Unscrew the protection cap, using the tool provided in the maintenance kit.
CAUTION:
Never remove the protection cap, unless you plan to replace the membrane.
Pull up the membrane holding ring with the tool provided in the maintenance kit.
Remove the old membrane and discard it.
6.3.2 Installing the Membrane
To install the new membrane, follow the steps below:
Note:
The membrane mounting surface must be clean and even (on top of the sensor, where the membrane and sensor have contact).
Replace the membrane
Note:
The 29039.0 Nitril
In the maintenance kit, pick up the two part membrane mounting tool.
Install the sleeve over the sensor head (end with shoulder downwards).
Note:
Once installed, a membrane cannot be reused. Avoid touching the membrane with bare fingers, as this may affect its sensitivity.