Hach 3654 Portable H2/N2 Analyzer manual Calibrating In a Liquid H2 or N2 Solution

Page 46

40 of 66

Calibrations - 3654 Portable H2/N2 Analyzer

 

 

4.2.4 Calibrating In a Liquid H2 or N2 Solution

The sensor can be calibrated in a standard liquid solution that has a known concentration of H2 or N2. You can create a solution with a known concentration by dissolving H2 or N2 in water at a measured temperature and pressure. Alternatively, analyze a carbonated beverage independently by some acceptable means.

1) Set the gas measurement phase to Dissolved and set the calibration medium to In a liquid or a gaseous sample at known concentration (as detailed in “Calibration Medium” on page 32).

2) Press the CAL button. Remember, this button may have been locked out to prevent an accidental reset (see “Locking Out the Instrument’s CAL Button” on page 32 for details).

3) A brief clearing [---]message appears.

4) Press CAL again within a 3 second period. The instrument then displays the concentration of H2 or N2, based on the last value of the calibration coefficient.

5) Modify this reading using the Up/Down Arrow keys until the displayed concentration agrees with that of the calibration solution.

6) Press CAL again.

7) Start the flow of the standard (calibration) solution through the flow chamber. Adjust the flow until the liquid is foam-free. The flow rate should be a minimum of 150 ml/ min and be stable.

Note:

A flow rate lower that 150 ml/min can be used for the calibration, but you must ensure that all measurements made after the calibration are made at exactly the same flow rate as is used for the calibration. To ensure accurate results it is recommended to use a model 32311 Flow Meter as described in “Model 32311 Flow Meter” on page 41.

If the instrument makes two consecutive measurements with less than 1% difference, it calibrates against this stabilized value. The LCD then displays the message [CAL] to indicate calibration has been successfully completed

If the calibration does not complete, the LCD displays the message [Err]. The reason for this calibration error is either that the measured gas partial pressure is under 5 mbar, or that a wrong instrument key was pressed during the calibration steps.

ORBISPHERE

Operator Manual

Image 46
Contents Operator Manual Page Product Recycling Information Orbisphere Product Disposal Following only applies to European customersBoard Analog Board Spacer External Connector Restriction of Hazardous SubstancesTable of Contents Maintenance and Troubleshooting Accessories and AttachmentsCalibrations Part Lists SpecificationsAppendix a Glossary Table of Contents 3654 Portable H2/N2 Analyzer About this Manual Safety ConventionsManual Overview Revision HistorySafety Recommendations Safety PrecautionsService and Repairs Precautionary Labels AcknowledgementsOrbisphere Installation What you have ReceivedSensor Installation WIN3654 PC Program InstallationAtmospheric Pressure Equilibrium Connections External Power Supply optionalPurge Gas Connection Instrument PC ConnectionPin RS232 Signal Use Flow Chamber Installation Installation Completion Check List Flow Chamber Dissolved Gas MeasurementsOrbisphere Operating Instructions Operating ControlsMeasurement Phase on Measurement Mode Select Gas Measurement PhaseTaking Measurements Storing Measurements In the Instrument Automatic Data AcquisitionManual Data Acquisition Viewing Stored Measurements Storing and Accessing Measurements From a ComputerDownloading Stored Values Copying Values Altering the Sampling Point DescriptionsSaving Values Printing Values Clearing Stored ValuesMonitoring Measurements In Real-Time Timebase Updating Rate Maximum Samples 10 divisionsAfter Use and Storage Options Setup Main Menu BasicsOptions Setup 3654 Portable H2/N2 Analyzer Gas Measurement Phase Reviewing Instrument ConfigurationConfiguring the Instrument Analyzer PC ConnectionMeasurement Units Automatic Data Acquisition Setting Sampling Intervals Membrane SelectionLocking Out the Instrument’s CAL Button Calibration MediumRolling Average Automatic Shutdown Orbisphere Calibrations Barometric Pressure Sensor CalibrationSensor Calibration Calibration Method Time Benefits Drawbacks ReferenceCalibration Setup Calibration In H2 or N2 at Atmospheric Pressure Calibration In H2 or N2 at Elevated Pressure Calibrating In a Liquid H2 or N2 Solution Accessories and Attachments Model 32311 Flow MeterModel 32939 External Power Supply Sensor Maintenance Maintenance and TroubleshootingPower Supply Maintenance ScheduleMounting Dual Membranes Membrane ReplacementRemoving the Membrane Installing the Membrane Orbisphere Place the membrane holding ring on the installation tool tip Purge Gas Cylinder Refill 1 N2 Purge Gas2 CO2 Purge Gas Filling the Purge Gas CylinderSerial Test Troubleshooting InstrumentKeyboard Test Clock Settings Display TestAnalog Voltages View Measurements ViewProgram Identification Troubleshooting OperationDiagnostics Messages Continuous PurgeTroubleshooting Table Symptom Cause Possible SolutionSpecifications Instrument SpecificationsSensor Specifications Sensor Type 31 270 HSensor Type 31 272 H 31 570 N PFA Etfe EctfeSpecifications 3654 Portable H2/N2 Analyzer Part Lists Instrument ConfigurationsTC Sensor and Parts on Configured System Analyzer Spare PartsSensor Spare Parts Terms and Definitions Appendix a GlossaryCommon Units UnitGlossary 3654 Portable H2/N2 Analyzer Annex Tables and IllustrationsAnnex 3654 Portable H2/N2 Analyzer User Notes