Hach 3654 Portable H2/N2 Analyzer manual Calibration In H2 or N2 at Atmospheric Pressure

Page 44

38 of 66

Calibrations - 3654 Portable H2/N2 Analyzer

 

 

4.2.2 Calibration In H2 or N2 at Atmospheric Pressure

The sensor can be calibrated in pure H2 or N2 gas at atmospheric pressure.

WARNING

Handle H2 gas with great care. It is extremely flammable and explosive.

To ensure that the flow chamber and sensor components are dry, remove and blow dry the flow chamber. Dry the sensor head surface with a clean soft tissue.

The analyzer uses an internal barometric pressure sensor during this calibration. First check that the pressure sensor is correct and does not need recalibrating. If it does need to be recalibrated then follow the instructions in “Barometric Pressure Sensor Calibration” on page 35 before attempting this calibration.

1)Set the gas measurement phase to Gaseous and set the calibration medium to In measured pure gas at atmospheric pressure (as detailed in “Calibration Medium” on page 32).

2)Connect a source of pure H2 or N2 gas to the inlet of the flow chamber, as illustrated in Fig 4-2 on page 37, and adjust the gas flow to one bubble per second when the exit tube is immersed in water.

3)Press the CAL button. Remember, this button may have been locked out to prevent an accidental reset (see “Locking Out the Instrument’s CAL Button” on page 32 for details).

4)Press the CAL button. Remember, this button may have been locked out to prevent an accidental reset (see “Locking Out the Instrument’s CAL Button” on page 32 for details).

5)A brief clearing message [---]appears.

6)Press CAL again within a 3 second period.

If the instrument makes two consecutive measurements with less than 1% difference, it calibrates against this stabilized value. The LCD then displays the message [CAL] to indicate calibration has been successfully completed

If the calibration does not complete, the LCD displays the message [Err]. The reason for this calibration error is either that the measured gas partial pressure is under 5 mbar, or that a wrong instrument key was pressed during the calibration steps.

ORBISPHERE

Operator Manual

Image 44
Contents Operator Manual Page Product Recycling Information Orbisphere Product Disposal Following only applies to European customersBoard Analog Board Spacer External Connector Restriction of Hazardous SubstancesTable of Contents Calibrations Accessories and AttachmentsMaintenance and Troubleshooting Appendix a Glossary SpecificationsPart Lists Table of Contents 3654 Portable H2/N2 Analyzer Safety Conventions Manual OverviewAbout this Manual Revision HistoryService and Repairs Safety PrecautionsSafety Recommendations Precautionary Labels AcknowledgementsOrbisphere Installation What you have ReceivedAtmospheric Pressure Equilibrium WIN3654 PC Program InstallationSensor Installation External Power Supply optional Purge Gas ConnectionConnections Instrument PC ConnectionPin RS232 Signal Use Flow Chamber Installation Installation Completion Check List Flow Chamber Dissolved Gas MeasurementsOrbisphere Operating Instructions Operating ControlsMeasurement Phase on Taking Measurements Select Gas Measurement PhaseMeasurement Mode Storing Measurements In the Instrument Automatic Data AcquisitionManual Data Acquisition Downloading Stored Values Storing and Accessing Measurements From a ComputerViewing Stored Measurements Saving Values Altering the Sampling Point DescriptionsCopying Values Printing Values Clearing Stored ValuesMonitoring Measurements In Real-Time Timebase Updating Rate Maximum Samples 10 divisionsAfter Use and Storage Options Setup Main Menu BasicsOptions Setup 3654 Portable H2/N2 Analyzer Reviewing Instrument Configuration Configuring the InstrumentGas Measurement Phase Analyzer PC ConnectionMeasurement Units Automatic Data Acquisition Setting Sampling Intervals Membrane SelectionRolling Average Calibration MediumLocking Out the Instrument’s CAL Button Automatic Shutdown Orbisphere Calibrations Barometric Pressure Sensor CalibrationSensor Calibration Calibration Method Time Benefits Drawbacks ReferenceCalibration Setup Calibration In H2 or N2 at Atmospheric Pressure Calibration In H2 or N2 at Elevated Pressure Calibrating In a Liquid H2 or N2 Solution Accessories and Attachments Model 32311 Flow MeterModel 32939 External Power Supply Maintenance and Troubleshooting Power SupplySensor Maintenance Maintenance ScheduleRemoving the Membrane Membrane ReplacementMounting Dual Membranes Installing the Membrane Orbisphere Place the membrane holding ring on the installation tool tip Purge Gas Cylinder Refill 1 N2 Purge Gas2 CO2 Purge Gas Filling the Purge Gas CylinderKeyboard Test Troubleshooting InstrumentSerial Test Clock Settings Display TestAnalog Voltages View Measurements ViewTroubleshooting Operation Diagnostics MessagesProgram Identification Continuous PurgeTroubleshooting Table Symptom Cause Possible SolutionSpecifications Instrument SpecificationsSensor Specifications Sensor Type 31 270 HSensor Type 31 272 H 31 570 N PFA Etfe EctfeSpecifications 3654 Portable H2/N2 Analyzer Instrument Configurations TC Sensor and Parts on Configured SystemPart Lists Analyzer Spare PartsSensor Spare Parts Appendix a Glossary Common UnitsTerms and Definitions UnitGlossary 3654 Portable H2/N2 Analyzer Annex Tables and IllustrationsAnnex 3654 Portable H2/N2 Analyzer User Notes