Hach 3654 Portable H2/N2 Analyzer manual Sensor Calibration

Page 42

36 of 66

Calibrations - 3654 Portable H2/N2 Analyzer

 

 

4.2 Sensor Calibration

The TC sensor can be calibrated in pure H2 or N2 gas at atmospheric or elevated pressure. The following table gives information regarding the different methods.

Table 4-1:Sensor Calibration Information

Calibration Method

Time

Benefits

Drawbacks

Reference

Gas at atmospheric

 

 

Less accurate than

Gas with known

5 min

Quick and easy

purity, certified

pressure

other methods

 

 

barometer

 

 

 

 

 

 

 

 

 

 

 

Higher accuracy than

 

 

Gas at elevated

10 min

atmospheric method.

Requires a certified

Certified pressure

pressure (2-3 bar)

Validation of the

pressure gauge

gauge

 

 

sensor linearity

 

 

 

 

 

 

 

Before calibrating check that the sensor membrane surface is tight, smooth and wrinkle free. Select the gas measurement phase for calibration (for full details see “Gas Measurement Phase” on page 29). Next, select the calibration medium using the WIN3654 program (see “Calibration Medium” on page 32).

Ensure that normal measurement mode is selected in the Measurement Mode menu of the WIN3654 program (see “Measurement Mode” on page 33).

Calibration must only be carried out once the TC sensor is giving a stable measurement. Expose the membrane to the calibration sample until this stable reading is obtained. This usually takes about 5 minutes.

For calibration it is important that the sensor's membrane is at the same temperature as the calibration sample. Therefore, allow some sample to pass through the flow chamber for 3 minutes before calibrating.

ORBISPHERE

Operator Manual

Image 42
Contents Operator Manual Page Product Recycling Information Orbisphere Product Disposal Following only applies to European customersBoard Analog Board Spacer External Connector Restriction of Hazardous SubstancesTable of Contents Accessories and Attachments Maintenance and TroubleshootingCalibrations Specifications Part ListsAppendix a Glossary Table of Contents 3654 Portable H2/N2 Analyzer About this Manual Safety ConventionsManual Overview Revision HistorySafety Precautions Safety RecommendationsService and Repairs Precautionary Labels AcknowledgementsOrbisphere Installation What you have ReceivedWIN3654 PC Program Installation Sensor InstallationAtmospheric Pressure Equilibrium Connections External Power Supply optionalPurge Gas Connection Instrument PC ConnectionPin RS232 Signal Use Flow Chamber Installation Installation Completion Check List Flow Chamber Dissolved Gas MeasurementsOrbisphere Operating Instructions Operating ControlsMeasurement Phase on Select Gas Measurement Phase Measurement ModeTaking Measurements Storing Measurements In the Instrument Automatic Data AcquisitionManual Data Acquisition Storing and Accessing Measurements From a Computer Viewing Stored MeasurementsDownloading Stored Values Altering the Sampling Point Descriptions Copying ValuesSaving Values Printing Values Clearing Stored ValuesMonitoring Measurements In Real-Time Timebase Updating Rate Maximum Samples 10 divisionsAfter Use and Storage Options Setup Main Menu BasicsOptions Setup 3654 Portable H2/N2 Analyzer Gas Measurement Phase Reviewing Instrument ConfigurationConfiguring the Instrument Analyzer PC ConnectionMeasurement Units Automatic Data Acquisition Setting Sampling Intervals Membrane SelectionCalibration Medium Locking Out the Instrument’s CAL ButtonRolling Average Automatic Shutdown Orbisphere Calibrations Barometric Pressure Sensor CalibrationSensor Calibration Calibration Method Time Benefits Drawbacks ReferenceCalibration Setup Calibration In H2 or N2 at Atmospheric Pressure Calibration In H2 or N2 at Elevated Pressure Calibrating In a Liquid H2 or N2 Solution Accessories and Attachments Model 32311 Flow MeterModel 32939 External Power Supply Sensor Maintenance Maintenance and TroubleshootingPower Supply Maintenance ScheduleMembrane Replacement Mounting Dual MembranesRemoving the Membrane Installing the Membrane Orbisphere Place the membrane holding ring on the installation tool tip Purge Gas Cylinder Refill 1 N2 Purge Gas2 CO2 Purge Gas Filling the Purge Gas CylinderTroubleshooting Instrument Serial TestKeyboard Test Clock Settings Display TestAnalog Voltages View Measurements ViewProgram Identification Troubleshooting OperationDiagnostics Messages Continuous PurgeTroubleshooting Table Symptom Cause Possible SolutionSpecifications Instrument SpecificationsSensor Specifications Sensor Type 31 270 HSensor Type 31 272 H 31 570 N PFA Etfe EctfeSpecifications 3654 Portable H2/N2 Analyzer Part Lists Instrument ConfigurationsTC Sensor and Parts on Configured System Analyzer Spare PartsSensor Spare Parts Terms and Definitions Appendix a GlossaryCommon Units UnitGlossary 3654 Portable H2/N2 Analyzer Annex Tables and IllustrationsAnnex 3654 Portable H2/N2 Analyzer User Notes