Hach 3654 Portable H2/N2 Analyzer Analyzer PC Connection, Reviewing Instrument Configuration

Page 35

3654 Portable H2/N2 Analyzer - Options Setup

29 of 66

 

 

3.2 Analyzer - PC Connection

For the hardware connection of the analyzer to a PC, see “Instrument - PC Connection” on page 11.

The Configuration, Serial port menu lets you choose one of four serial communication ports, as follows:

Usually, COM1 is used to connect to a mouse, so try COM2 first. You may find that a separate program supplied with your PC may be necessary to activate this port.

Click on OK to activate the selected port. If the port you have selected here is available, the WIN3654 program will return to the main menu. Otherwise, you will see an RS232 ERRORS message advising you to select another port.

3.3 Reviewing Instrument Configuration

To review if the analyzer is set up as expected, choose the Configuration, Configuration view command. You will see a window similar to the following:

Some of the parameters listed, can be changed by using the WIN3654 program.

However, should you see any unexpected items listed on your screen that you cannot correct, contact your Hach Ultra representative.

3.4 Configuring the Instrument

The ORBISPHERE 3654 analyzer can be readily configured for your application using the following commands in the Configuration menu. The instrument must be connected to your PC and powered on, in order to change any configuration parameters.

3.4.1 Gas Measurement Phase

The instrument can measure N2 or H2 in either a liquid or a gaseous sample, but you must select which gas phase to use.

Choosing the WIN3654 program's Configuration, Gas Phase menu lets you select the gas measurement phase.

Operator Manual

ORBISPHERE

Image 35
Contents Operator Manual Page Product Recycling Information Orbisphere Following only applies to European customers Product DisposalRestriction of Hazardous Substances Board Analog Board Spacer External ConnectorTable of Contents Calibrations Accessories and AttachmentsMaintenance and Troubleshooting Appendix a Glossary SpecificationsPart Lists Table of Contents 3654 Portable H2/N2 Analyzer Revision History Safety ConventionsManual Overview About this ManualService and Repairs Safety PrecautionsSafety Recommendations Acknowledgements Precautionary LabelsOrbisphere What you have Received InstallationAtmospheric Pressure Equilibrium WIN3654 PC Program InstallationSensor Installation Instrument PC Connection External Power Supply optionalPurge Gas Connection ConnectionsPin RS232 Signal Use Flow Chamber Installation Installation Completion Check List Dissolved Gas Measurements Flow ChamberOrbisphere Operating Controls Operating InstructionsMeasurement Phase on Taking Measurements Select Gas Measurement PhaseMeasurement Mode Automatic Data Acquisition Storing Measurements In the InstrumentManual Data Acquisition Downloading Stored Values Storing and Accessing Measurements From a ComputerViewing Stored Measurements Saving Values Altering the Sampling Point DescriptionsCopying Values Clearing Stored Values Printing ValuesTimebase Updating Rate Maximum Samples 10 divisions Monitoring Measurements In Real-TimeAfter Use and Storage Main Menu Basics Options SetupOptions Setup 3654 Portable H2/N2 Analyzer Analyzer PC Connection Reviewing Instrument ConfigurationConfiguring the Instrument Gas Measurement PhaseMeasurement Units Membrane Selection Automatic Data Acquisition Setting Sampling IntervalsRolling Average Calibration MediumLocking Out the Instrument’s CAL Button Automatic Shutdown Orbisphere Barometric Pressure Sensor Calibration CalibrationsCalibration Method Time Benefits Drawbacks Reference Sensor CalibrationCalibration Setup Calibration In H2 or N2 at Atmospheric Pressure Calibration In H2 or N2 at Elevated Pressure Calibrating In a Liquid H2 or N2 Solution Model 32311 Flow Meter Accessories and AttachmentsModel 32939 External Power Supply Maintenance Schedule Maintenance and TroubleshootingPower Supply Sensor MaintenanceRemoving the Membrane Membrane ReplacementMounting Dual Membranes Installing the Membrane Orbisphere Place the membrane holding ring on the installation tool tip 1 N2 Purge Gas Purge Gas Cylinder RefillFilling the Purge Gas Cylinder 2 CO2 Purge GasKeyboard Test Troubleshooting InstrumentSerial Test Display Test Clock SettingsMeasurements View Analog Voltages ViewContinuous Purge Troubleshooting OperationDiagnostics Messages Program IdentificationSymptom Cause Possible Solution Troubleshooting TableInstrument Specifications SpecificationsSensor Type 31 270 H Sensor SpecificationsPFA Etfe Ectfe Sensor Type 31 272 H 31 570 NSpecifications 3654 Portable H2/N2 Analyzer Analyzer Spare Parts Instrument ConfigurationsTC Sensor and Parts on Configured System Part ListsSensor Spare Parts Unit Appendix a GlossaryCommon Units Terms and DefinitionsGlossary 3654 Portable H2/N2 Analyzer Tables and Illustrations AnnexAnnex 3654 Portable H2/N2 Analyzer User Notes