Hach 3654 Portable H2/N2 Analyzer manual Calibration In H2 or N2 at Elevated Pressure

Page 45

3654 Portable H2/N2 Analyzer - Calibrations

39 of 66

 

 

4.2.3 Calibration In H2 or N2 at Elevated Pressure

The sensor can be calibrated in pure H2 or N2 gas at an elevated pressure. This method requires an accurate pressure gauge connected to the exit of the flow chamber.

WARNING

Handle H2 gas with great care. It is extremely flammable and explosive.

When calibrating in gas phase, the flow chamber and sensor head must be dry.To ensure that the flow chamber and sensor components are dry, remove and dry them with a clean soft tissue.

1)Set the gas measurement phase to Gaseous and set the calibration medium to In a liquid or a gaseous sample at known concentration (as detailed in “Calibration Medium” on page 32).

2)Connect a source of pure H2 or N2 gas to the inlet of the flow chamber, as illustrated in Fig 4-2 on page 37, and adjust the gas flow exiting from the flow chamber to be in the range of 1 to 5 bar (it is best to use a pressure close to the application conditions).

3)Press the CAL button. Remember, this button may have been locked out to prevent an accidental reset (see “Locking Out the Instrument’s CAL Button” on page 32 for details).

4)A brief clearing message [---]appears.

5)Press CAL again within a 3 second period. The instrument then displays the absolute pressure of calibration gas (i.e. gauge pressure plus atmospheric pressure).

6)Modify this reading using the Up/Down Arrow keys until the displayed pressure agrees with that of the gauge plus atmospheric pressure.

7)Press CAL again.

If the instrument makes two consecutive measurements with less than 1% difference, it calibrates against this stabilized value. The LCD then displays the message [CAL] to indicate calibration has been successfully completed

If the calibration does not complete, the LCD displays the message [Err]. The reason for this calibration error is either that the measured gas partial pressure is under 5 mbar, or that a wrong instrument key was pressed during the calibration steps.

Operator Manual

ORBISPHERE

Image 45
Contents Operator Manual Page Product Recycling Information Orbisphere Following only applies to European customers Product DisposalRestriction of Hazardous Substances Board Analog Board Spacer External ConnectorTable of Contents Accessories and Attachments Maintenance and TroubleshootingCalibrations Specifications Part ListsAppendix a Glossary Table of Contents 3654 Portable H2/N2 Analyzer Manual Overview Safety ConventionsAbout this Manual Revision HistorySafety Precautions Safety RecommendationsService and Repairs Acknowledgements Precautionary LabelsOrbisphere What you have Received InstallationWIN3654 PC Program Installation Sensor InstallationAtmospheric Pressure Equilibrium Purge Gas Connection External Power Supply optionalConnections Instrument PC ConnectionPin RS232 Signal Use Flow Chamber Installation Installation Completion Check List Dissolved Gas Measurements Flow ChamberOrbisphere Operating Controls Operating InstructionsMeasurement Phase on Select Gas Measurement Phase Measurement ModeTaking Measurements Automatic Data Acquisition Storing Measurements In the InstrumentManual Data Acquisition Storing and Accessing Measurements From a Computer Viewing Stored MeasurementsDownloading Stored Values Altering the Sampling Point Descriptions Copying ValuesSaving Values Clearing Stored Values Printing ValuesTimebase Updating Rate Maximum Samples 10 divisions Monitoring Measurements In Real-TimeAfter Use and Storage Main Menu Basics Options SetupOptions Setup 3654 Portable H2/N2 Analyzer Configuring the Instrument Reviewing Instrument ConfigurationGas Measurement Phase Analyzer PC ConnectionMeasurement Units Membrane Selection Automatic Data Acquisition Setting Sampling IntervalsCalibration Medium Locking Out the Instrument’s CAL ButtonRolling Average Automatic Shutdown Orbisphere Barometric Pressure Sensor Calibration CalibrationsCalibration Method Time Benefits Drawbacks Reference Sensor CalibrationCalibration Setup Calibration In H2 or N2 at Atmospheric Pressure Calibration In H2 or N2 at Elevated Pressure Calibrating In a Liquid H2 or N2 Solution Model 32311 Flow Meter Accessories and AttachmentsModel 32939 External Power Supply Power Supply Maintenance and TroubleshootingSensor Maintenance Maintenance ScheduleMembrane Replacement Mounting Dual MembranesRemoving the Membrane Installing the Membrane Orbisphere Place the membrane holding ring on the installation tool tip 1 N2 Purge Gas Purge Gas Cylinder RefillFilling the Purge Gas Cylinder 2 CO2 Purge GasTroubleshooting Instrument Serial TestKeyboard Test Display Test Clock SettingsMeasurements View Analog Voltages ViewDiagnostics Messages Troubleshooting OperationProgram Identification Continuous PurgeSymptom Cause Possible Solution Troubleshooting TableInstrument Specifications SpecificationsSensor Type 31 270 H Sensor SpecificationsPFA Etfe Ectfe Sensor Type 31 272 H 31 570 NSpecifications 3654 Portable H2/N2 Analyzer TC Sensor and Parts on Configured System Instrument ConfigurationsPart Lists Analyzer Spare PartsSensor Spare Parts Common Units Appendix a GlossaryTerms and Definitions UnitGlossary 3654 Portable H2/N2 Analyzer Tables and Illustrations AnnexAnnex 3654 Portable H2/N2 Analyzer User Notes